The System Diagnostics menu can be used to perform functional
diagnostic tests. Use the following procedure to access the System
Diagnostics menu.
1. Power on the MAC 800 system by pressing the Power button.
2. From the Main Menu, press F4 to select System Configuration.
3. Press F6 (More) > F6 (More) > F5 (Service Setup).
The following window prompts you to enter the Service password.
4. Type 7763 and press F6 to select OK to open the Service Setup menu.
5. Move the focus to the System Diagnostics button and press the Enter
key to open the Diagnostic Tests window.