Rs232 Test - GE MAC 800 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 800:
Table of Contents

Advertisement

RS232 Test

2031504-159D
5. Perform the Print Head Test.
a. Select the Print Head Test button.
The writer prints a 1-page print head test report.
b. Verify that there are no gaps in any of the lines printed.
Up to 5 mm of blank paper is allowable at the top and bottom of
the page.
When the page is done printing, the following window opens.
c.
Do one of the following:
If there are no gaps in the lines on the printed report, press
F4 to select Yes.
If there are gaps in the lines on the printed report, press F5
to select No.
Replace the printer as described in
Assembly"
6. When all writer tests have been performed, press F6 to select Cancel
and close the window.
The RS232 Test can be used to determine if the comm ports are working
properly.
1. Open the Diagnostic Tests window as described in
System Diagnostics Function"
2. Use a paper clip to short pins 2 and 3 in the COM port.
3. Select the RS232 Test button.
The following window opens.
MAC™ 800
on page 4-13.
on page 3-6
Troubleshooting
"Replacing the Printer
"Accessing the
3-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents