2 Setup methods according to ISO 17664 | 2.7 Automatic cleaning programmes
fresh, clear water (limited rinsing). The water is aspirated from the Aquamat
through the filter housing to the separation system. The spittoon bowl rinsing is
alternately turned on and off.
The "After treatment" cleaning programme prevents the coagulation and adhe-
sion of proteins inside the drainage system and improves the removal of fine
particles. In addition, the cleaning programme includes automatic rinsing of the
water-conducting system and all instruments for 20 seconds each including
tumbler filling.
▶ Run after each treatment.
ð
The following message is displayed after completion of the cleaning pro-
gramme:

2.7.4 Cleaning programs "Morning" and "Evening"

The "Morning" and "Evening" cleaning programmes are automatic programmes
for hydro-mechanical cleaning of the suction hoses, drainage and separation
systems. The cleaning process is carried out by supplying fresh, clear water and
cleaning agent. The water is sucked out of the Aquamat through the extraction
point and suction handpiece to the separation system. The spittoon bowl rinsing
is alternately turned on and off.
The "Morning" cleaning programme includes automatic rinsing of the water-
conducting system and all instruments for 2 minutes each including tumbler
filling. In addition, the suction system is rinsed with water.
In the short cleaning programme "Evening", all instruments on the dentist ele-
ment and assistant element are rinsed for 20 seconds each. In addition, the
suction system and the spittoon bowl are cleaned with DEKAPURE Gel. KaVo re-
commends running the "Evening" cleaning programme after surgical interven-
tions.
The as DEKAPURE Gel is automatically suctioned. As soon as the DEKAPURE Gel
bottle is empty, the display on the dentist element shows a status message.
Once the message " Replace DEKAPURE Gel bottle" is displayed, a full
DEKAPURE Gel bottle must be inserted.
Note
If no DEKAPURE Gel is used, DEKAMAT has to be deactivated in the menu
"settings".