Reprocessing Of The Swing-Arm S Table (E70 Vision Only) - KaVo ESTETICA E70 Vision Care Instructions

Hide thumbs Also See for ESTETICA E70 Vision:
Table of Contents

Advertisement

2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
Sterilize the instrument holder
Damage caused by sterilisation.
▶ Do not sterilise the instrument holder.
2.4.3 Reprocessing of the swing-arm S table (E70 Vision
only)
Note
The swing-arm S-table (Mat. no. 10074755)) can be removed to facilitate
cleaning.
Cleaning and disinfection of the instrument arm S table
Damage caused by treatment in a washer disinfector.
▶ Do not sterilise the instrument arm S table in a washer disinfector.
Note
No cleaning programmes can be run when the swing-arm is removed.
▶ Locking the screen.
See also:
2 Locking the touchscreen, Page 9
▶ Take the handpieces off the hoses.
▶ Take the roller out of the swing-arm and then place the hose in the hygiene
centre.
▶ Pull-off the swing-arm in an upward direction.
▶ Clean the swing-arm and roller under running water.
Care instructions ESTETICA E70 Vision / E80 Vision