Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
Storage of the assembly kit for physiological saline
solution
Observe all necessary measures for hygiene when storing sterile goods. Store
protected from dust and in a dry place, release with identification on the pack-
aging. Evaluate the duration of storage.

2.4.8 Reprocessing KaVo ERGOcam One

Note
The camera handle cannot be sterilised or disinfected with heat.
Cleaning
Note
Disinfect the camera before cleaning.
Disinfect after cleaning if necessary.
Manual Exterior Cleaning
Damage to the camera due to improper cleaning.
Destroys the surface or the optical system of the camera.
▶ Do not use abrasives or aggressive cleansers!
▶ Do not immerse the camera in liquids!
▶ Switch main device switch off.
▶ Remove the camera.
▶ Clean the surface with a soft cloth and mild liquid cleaning agent.
Disinfection
Manual external disinfection
Highly flammable gases-.
Explosion.
▶ Allow product to cool before disinfecting.
NOTICE
WARNING