9.2.2 Adjustment of temperature value

The temperature is measured at two selectable points and it is possible to correct it
there.
 Invoke service function „Tmp Adj", see chap. 9.2
 Confirm by pressing the PRINT button. The current setting for the first
temperature point is displayed. Or leave it on factory setting 80°C or select a
desired temperature with the help of the navigation buttons   (available 50-
130°C).
 Confirm with the PRINT button, the first heat-up phase is started. The remaining
time is displayed.
 Temperature calibration for the first point takes 15 min. Compare the temperature
displayed on the DLB-A01N with that of the moisture analyzer. If the two values
do not match, correct them using the navigation buttons   and confirm with the
PRINT button
 The current setting for the second temperature point is displayed. Or leave it on
factory setting 150°C or select the desired temperature with the help of the
navigation buttons  . The second temperature point has to be at least 30°C
over the first one, max. 160°C.
 Confirm with the PRINT button, the second heat-up phase is started. The
remaining time is displayed.
 Temperature calibration for the second point takes 15 min. Compare the
temperature displayed on the DLB-A01N with that of the moisture analyzer. If the
two values do not match, correct with the help of the navigation buttons  .
 Confirm by pressing the PRINT button. The adjustment is completed, the
appliance returns into the menu. Use the ON/OFF button to return into
22
.
weighing mode.