KERN DLB 160-3A Operating Instructions Manual page 99

Electronic moisture analyser
Hide thumbs Also See for DLB 160-3A:
Table of Contents

Advertisement

Available languages

Available languages

Start of moisture analysis:
Preheating stage enabled
 Confirm by the PRINT button, close
the sample chamber and wait for the
warm-up phase.
 After reaching the set temperature
„ready" will be displayed.
 Confirm by pressing the PRINT
button. The weight display and the
indicator „" are displayed.
 If required, tare by using TARE
button.
 Put prepared sample (see chap. 13.4)
in the sample dish and close the
sample chamber.
 Wait for stability display, then press
the PRINT button. Drying is started.
The result display appears.
The indicator „H" displays the active
drying process.
DLB_A-BA-e-1714
Preheating stage disabled
 Confirm by pressing the PRINT
button. The weight display and the
indicator „" are displayed.
 If required, tare by using TARE
button.
 Put prepared sample (see chap. 13.4)
in the sample dish and close the
sample chamber.
 Wait for stability display, then press
the PRINT button. Drying is started.
The result display appears.
The indicator „H" displays the active
drying process.
43

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dlb-a series