Start of moisture analysis:
Preheating stage enabled
 Confirm by the PRINT button, close
the sample chamber and wait for the
warm-up phase.
 After reaching the set temperature
„ready" will be displayed.
 Confirm by pressing the PRINT
button. The weight display and the
indicator „" are displayed.
 If required, tare by using TARE
button.
 Put prepared sample (see chap. 13.4)
in the sample dish and close the
sample chamber.
 Wait for stability display, then press
the PRINT button. Drying is started.
The result display appears.
The indicator „H" displays the active
drying process.
36
Preheating stage disabled
 Confirm by pressing the PRINT
button. The weight display and the
indicator „" are displayed.
 If required, tare by using TARE
button.
 Put prepared sample (see chap. 13.4)
in the sample dish and close the
sample chamber.
 Wait for stability display, then press
the PRINT button. Drying is started.
The result display appears.
The indicator „H" displays the active
drying process.