7.2 List of status messages

Message
Description, process
a) Faults, alarms:
Restart
If this message appears whilst in operation, the device is faulty.
Please contact Elster support.
C-fact. err.
The conversion factor could not be calculated as the measured gas temperature
was outside -100°C to +100°C or no usable compressibility ratio factor was
available.
Data error
An error was detected during the cyclic check of the data subject to calibration
regulations. Please contact Elster support.
T Alarm Lim.
The gas temperature measured is outside acceptable limits.
The limit values can be changed when putting the device into operation in the
presence of a calibration officer (see Operating Instructions [1]).
p Alarm Lim.
The measured gas temperature is outside the permissible limits.
The limit values can be changed when putting the device into operation in the
presence of a calibration officer (see Operating Instructions [1]).
K-val. error
The compressibility ratio factor could not be calculated as no valid real gas factor
could be determined.
z-fact. err.
The real gas factor could not be calculated.
At least one of the gas analysis values Ho.n, CO2, H2, rhon, is outside the
permissible range.
No volume was recorded at the used input for conversion. (The automatic switchover of the
Vm-Alarm
pulse transducer to a second pulse transducer is not configured acc. chapter 4.4, page 15.
T Inp. error
The gas temperature cannot be measured due to a fault.
Please contact Elster support.
p Inp. error
The gas pressure cannot be measured due to a fault.
Please contact Elster support.
b) Warnings:
Dat.restaur.
E The batteries of the EK280 are intermittently dropping out. As a result of this, the
time has not changed and no measurement and volume conversion have happened.
However, all data is available.
This message appears when old batteries are removed before connecting the new
ones (see Operating Instructions [1]).
Proceed with the help of the configuration program "enSuite":
Outp.1 Error
There should be more pulses being emitted from the specified output than are
permitted under its configured settings.
Outp.2 Error
In order to eliminate the cause of the problem, you can use the configuration
Outp.3 Error
program "enSuite" to either:
Outp.4 Error
In both cases, the change must be aligned with the connected device.
I2 Pulse cmp
In the pulse comparison, e.g. between input 1 and input 2, the deviation was too
high (see chapter 4.5, page 17).
W Warn Lim.
The displayed value is outside the set warning limit values.
The warning limits can be adjusted using the configuration program "enSuite".
Application Manual
Set the time of the EK280
Delete the status register
reduce the cp value of the output
or increase the output frequency.