7.9
Ramp function
With the ramp function, temperature changes over any time period can be conveniently entered. This is
especially advantageous with very low temperature changes (e.g. 0.1 °C/ day).
Example: From the current outflow temperature (e.g. 242.4 °C) 200 °C of cooling is to occur over 5
days.
Then 200 °C is entered as the temperature change, the time value 5 is entered for the time and day(s)
selected as the time unit.
The ramp function is executed until it is manually terminated or until the temperature limits
T
Command
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Help
Menu
Status
Temp. change
Time
Time unit
Help
Menu
70
(min) or T
(max) described in Section 7.6.2 are attained.
il
ih
Program 1
Program 2
Program 3
Program 4
Program 5
Ramp function
End
T
set
Second(s)
Minute(s)
Hour(s)
Day(s)
End
T
set
−
Ramp function
− Open the list of device parameters using
the soft key
− With the cursor keys continue to
− Confirm with the key
T
fix
− Enter a positive or negative temperature
value with Temp. change .
− With Time enter a figure (without time
unit).
− With Time unit choose between
Second(s)
− Under Status the ramp is started
− When the ramp function is being
executed,
window bar.
T
fix
− Without manual switch-off the ramp
terminates at the latest at T
(max).
Starting up
Proline Kryomats
Menu .
Programmer
Ramp function
.
up to
Day(s)
.
Start
or stopped
Stop
Ramp active
appears in the
YAUE0007 / 2/26/2009
.
.
(min) or T
il
ih