12 TECHNICAL DATA

12.1 MEASUREMENT SPEED
>60 readings per minute.
When measuring high temperature materials
measurement speed must be reduced to prevent
overheating
of
the
measurement speed of the High Temperature
PINIP™ probe at 250°C (480°F) is 4 readings per
minute.
12.2 MINIMUM SUBSTRATE THICKNESS
Ferrous: 300 µm (12 mils)
Non-ferrous: 100 µm (4 mils)
Measurements can be taken on thinner substrates
if 2-point calibration is carried out either side of the
required substrate thickness, however gauges will
have reduced range when adjusted for thin
substrates.
12.3 PROBE OPERATING TEMPERATURE
Separate ferrous probes:
High temperature PINIP™ probes: 250°C (480°F)
probe.
The
maximum
150°C (300°F)
Miniature probes without outer sleeve:
All other probes:
12.4 PHYSICAL
Weight
Separate Probe (FNF1), 190g
(including
(6.7oz)
batteries):
Separate Probe (PINIP™), 155g
(5.5oz)
Integral Probe, 130g (4.6oz)
Dimensions: 130 mm x 70 mm x 35 mm
(5.12" x 2.76" x 1.38")
Gauge
0°C to 50°C (32°F to 120°F)
operating
Operation outside these limits
temperature:
depends upon climatic conditions.
Case:
High impact ABS
12.5 POWER SUPPLY
Internal batteries, 2 x LR03 (AAA), alkaline
batteries or rechargeable
Battery life
n
30
hours to 40 hours continuous use with alkaline
dry batteries. (15 000 to 20 000 readings at an
150°C (300°F)
80°C (176°F)
l
m
equivalents.
R
dry
39