Calibrating High Temperature PINIP™ Probes
Special thickness standards are supplied with
F1 2 High Temperature PINIP™ Probes - see
page 35. These thickness standards should be
used in place of the calibration standard in Step 1
of "Calibration adjustment procedure" on page 24.
1. Place appropriate thickness standard over end
of PINIP™ probe.
2. Press CAL softkey.
3. Place probe on hot
reading.
4. Lift probe and then replace on hot surface to
take second reading.
5. Repeat as necessary until reading is stable.
6. Press SET to accept value.
7. Remove thickness standard from end of
PINIP™ probe.
8. Proceed with Step 2 - see page 26.
e.
The temperature of the surface used for calibration
should be equal to the temperature of the substrate
being measured.
e
surface and take a

6 STATISTICS

3
The Elcometer 456
Basic has a Simple Statistics
feature (STATS) which calculates and displays a
statistical analysis of readings as they are taken.
When EXTENDED MENU is active, press STATS
softkey to access STATS MENU (Figure 33).
Figure 33. Stats menu
The statistics available are:
Number of Readings
Mean
Standard Deviation
Coefficient of Variation
Highest Reading
Lowest Reading
% greater to or equal to NDFT
% between 90% and 100% of NDFT
90/10 pass or fail
R
29