reading. Repeat this action until a stable
reading is obtained.
First reading
Figure 27. Step 2 - Calibration adjustment on
uncoated sample
To reject the displayed reading and start Step
2 of the calibration procedure again, press the
Reset softkey
3. Press ZERO softkey to zero the display (Figure
28).
Figure 28. Zero the display
4. Press SET softkey to accept this value.
Second reading
.
The gauge will display the option to test the
calibration of the gauge.
Figure 29. TEST READINGS screen
5. Either press NO softkey to complete the
calibration adjustment procedure and return
the gauge to the reading screen, or proceed to
take test readings - see Taking test readings
below.