7 TECHNICAL SPECIFICATION

Measurement Principle
Measurement Mode
Minimum Sample Volume
Measurement Range
b
Conversion Factor
Resolution
Accuracy
Operating Temperature
Operating Humidity
Battery Type
Battery Life
Dimensions
Weight
b
2
Based on an area of 12.5cm and a sample volume of 3ml.
en-11
2 Electrode Bioplar AC
ISO, IMO, Temperature
ISO Mode
2
0 - 2399µg/cm
2
µS/cm to µg/cm : 0.12
2
µS/cm to mg/m : 1.2
2
0 - 239.9µg/cm : 0.1µg/cm
2
240 - 2399µg/cm : 1µg/cm
±2% of full scale (for each range)
5°C to 40°C (41°F to 105°F)
85% or less relative humidity (no condensation)
Approximately 200 hours continuous use
164 x 29 x 20mm (6.5 x 1.1 x 0.79")
50g (1.76 oz) - including sensor and batteries
0.12ml
0 - 2199µg/cm
µS/cm to µg/cm : 0.11
µS/cm to mg/m : 1.1
2
0 - 219.9µg/cm : 0.1µg/cm
2
220 - 2199µg/cm : 1µg/cm
2 x CR2032 lithium
without backlight
IMO Mode
2
2
2
2
2
2
2
www.elcometer.com