Installation Procedure

Analyzer Information

1. Connect a power cord to the analyzer and turn on the analyzer.
2. After the analyzer has completed turning on, press System, Show System. Make note of the
following information from the Show System screen:
Product Number ________________
Serial Number ________________
Instrument S/W Revision ________________
3. Check for the presence of the options listed below in the Show System screen. Put a check
mark or "X" after each option listed below that appears in the Show System menu.
N9020B-503 ______
N9020B-508 ______
N9020B-513 ______
N9020B-526 ______
N9020B-BBA ______
N9020B-B40 ______
N9020B-DP2 ______
N9020B-MPB ______
Refer to the following conditions:
— If the analyzer is not equipped with N9020B-B40, do not proceed with the installation of
this upgrade kit. If N9020B-B40 is not installed, other upgrade kits are available to upgrade
to 85 MHz, 125 MHz, or 160 MHz analysis bandwidth.
— If the analyzer has N9020B-BBA, do not proceed with the installation of this kit. N9020B
analyzers with Option BBA do not have sufficient room to install the two additional boards
required for analysis bandwidths of 85 MHz and wider.
— If the N9020B-MPB is not already present and either frequency range option N9020B-508,
N9020B-513, or N9020B-526 is present, it will be necessary to upgrade the analyzer to add
option N9020B-MPB before installing this kit. Refer to
http://www.keysight.com/find/mxa_upgrades
MPB.
Installation Note N9020-90254
Do not attempt to install this kit on an N9020A. Although much of the hardware is identical, the
licenses included in this kit will only work on an N9020B.
for details on how to upgrade to N9020B-