Tools Required; Initial Instrument Functionality Check - Keysight Technologies N9020B Installation Note

Mxa signal analyzer, option bu1/2/3, 40 mhz to 85/125/160 mhz analysis bandwidth upgrade
Table of Contents

Advertisement

Option BU1/BU2/BU3, 40 MHz to 85/125/160 MHz Analysis Bandwidth Upgrade

Tools Required

— T-10 TORX Driver
— T-20 TORX Driver
— Keysight Calibration and Adjustment Software, N7814A (revision E.16.00 or later)
— Test equipment and computer supported by the X-Series Performance Tests and Adjustment
Software
— MXA Signal Analyzer Service Guide. This manual is available online at the following URL:
www.keysight.com/find/N9020B_service_guide
— Microsoft Windows based personal computer with internet access and USB port
— USB storage device with >2GB free memory

Initial Instrument Functionality Check

Power on the instrument and allow the instrument to boot up. Run an alignment and display the
measurement screen. (The instrument will probably display a spectrum analyzer screen and you will
see the instrument sweeping.)
There should be no alignment failures. If there are failures, investigate and fix the problem before
continuing.
6
Before you disassemble the instrument, turn the power switch to Standby. After the instrument
has completely shut down, unplug the instrument. Failure to unplug the instrument can result in
personal injury.
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on
electronic assemblies should be performed at a static-safe workstation. Refer to the
documentation that pertains to your instrument for information about static-safe workstations
and ordering static-safe accessories.
Installation Note N9020-90254