Keysight Technologies N9020B Installation Note page 24

Mxa signal analyzer, option bu1/2/3, 40 mhz to 85/125/160 mhz analysis bandwidth upgrade
Table of Contents

Advertisement

Installation Procedure
10.Locate W53 in the Opt BU1/BU2/BU3 Cable Kit. This cable will have part number 8121-2292
and will have ends labeled "901" and "101". Connect the end labeled "901" to A15J901 on the
A15 Enhanced Front End Control Assembly. Route the cable along the right side chassis and
connect the other end to A25J101 on the A25 Wideband Analog IF Assembly.
11.Refer to
Figure
15. Dress the coaxial cables neatly and snap into the cable clip with the correct
sequence as shown in Figure 8.
Figure 15
Dressing Cables to A25 Wideband Analog IF
12.Refer to
Figure
16. Re-install the wire cable hold down. Assure the cables lay flat.
13.Attach two cable ties around the bundle of gray cables.
Figure 16
Cables Under Cable Hold Down
24
Installation Note N9020-90254

Advertisement

Table of Contents
loading

Table of Contents