4.3 Placing of probes and sensors

Warning:
The following rules must be followed precisely to avoid the problems described below:

4.3.1 Absolute rule

Use only the cables, sensors and cuffs supplied by SCHILLER.

4.3.2 Probes and sensors to be used and positioning instructions

4.3.2.1 SpO2 sensor

Caution:
Warning:
Part number: 0-48-0080
The recommendations relate to the type of sensors and probes to be
used, the positioning of the sensors and probes on the patient and the
placing of the cables that connect the sensors and probes on the
patient and the device.
1. Make sure that the Oximeter function is on and that the oximeter
parameters are set correctly.
Use only the sensors from the list of accessories supplied by
SCHILLER.
2. Connect the oximeter patient cable to the oximeter connector and
give the connector a quarter turn to the right-hand side to lock it in
place (to disconnect, turn the connector to the left and pull it out).
3. Place the sensor over the nail of the index finger or equivalent site;
place the cable on the top of the hand and fix it using Velcro™ strap
or secure to the patient's wrist without tightening it.
Be cautious in placing the cable correctly so as to prevent damage
or injury (tangling and/or strangulation).
Page 14