Installation
Power Supply
Power Supply
Before turning on the E5072A power, check the following.
Verification of the Power Supply
Confirm that the power supplied to the E5072A meets the following
requirements:
Voltage
Frequency
Maximum Power consumption
Verification and Connection of Power Cable
The three- wire power cable attached to the E5072A has one wire serving as a
ground. Using this power cable allows the E5072A to be grounded, thereby
protecting you against electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
NEVER use a power cable showing any sign of damage. Faulty cables can
cause electrical shock.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 5 on page 30) on the rear panel of the E5072A and a three- wire power
outlet with the grounding prong firmly connected in the ground slot.
Use the supplied power cable with grounding wire to securely ground the
E5072A.
Power cord information is available on
https://www.keysight.com/find/powercords.
Keysight E5072A Network Analyzer
1. Switched automatically by the E5072A in conformity to the voltage used.
Requirements
90 to 132 VAC or 198 to 264 VAC
47 to 63 Hz
350 VA
1