Environmental Requirements; Operating Environments; Ventilation Requirement - Keysight Technologies E5072A Installation Manual

Network analyzer
Hide thumbs Also See for E5072A:
Table of Contents

Advertisement

Installation

Environmental Requirements

3. A packaged item supplied with the analyzer is missing.
4. A fault has been detected in the subsequent operation check of the
If an abnormality is detected in step 1, contact the company that transported
the analyzer as well as your nearest Keysight Technologies sales and service
office. For inspection by the transport company, save the packing box, shock-
absorbing material, and packaged items as you received them.
Environmental Requirements
Set up the E5072A where the following environmental requirements are met.

Operating Environments

Ensure that the operating environment meets the following requirements.
Temperature
Temperature range at the
error-correction
Humidity
Altitude
Vibration
Above environmental requirements are NOT for the specifications and
measurement accuracy of the analyzer, but for the operating environment
of the analyzer.

Ventilation Requirement

To ensure safety requirements, the specifications and measurement accuracy
of the analyzer, you must keep the environmental temperature within the
specified range by providing appropriate cooling clearance around the analyzer
or, for the rack mount type, by forcefully air- cooling inside the rack housing.
For more information on environmental temperature to satisfy the
specifications and measurement accuracy of the analyzer, see the specification
in the E5072A Online Help.
When the environmental temperature around the analyzer is kept within the
temperature range of the operating environment specification (See Section
"Operating Environments" on page
14
analyzer.
5 ºC to 40 ºC
23 ºC ± 5 ºC (< 1 ºC deviation from the temperature
when performing the error-correction)
20% to 80% at wet bulb temperature < +29 °C
(non-condensation)
0 to 2,000 m (0 to 6,561 feet)
0.21 G maximum, 5 Hz to 500 Hz
14), the analyzer conforms to the
Keysight E5072A Network Analyzer

Advertisement

Table of Contents
loading

Table of Contents