Operation
Impedance Analyzer Setup
8. With nothing connected to the 3.5-mm port, press the Open [-] key to start
9. Connect the SHORT to the 3.5-mm port.
10.Press the Short [-] key to start the short data measurement. When the short
11.Remove the SHORT from the 3.5-mm port of the 42941A. Then connect the
12.Press the Load [-] key to start the load data measurement. When the load
13.Press the Save Impedance key.
14
open data measurement. When the open data measurement is completed,
the softkey label changes to Open [Done].
data measurement is completed, the softkey label changes to Short [Done].
LOAD to the 3.5-mm port.
data measurement is completed, the softkey label changes to Load [Done].
Keysight 42941A Impedance Probe Kit