Fixture Compensation

Fixture Compensation
In an actual measurement, a probe adapter is attached to the probe. Fixture
compensation is required for compensating residual impedance and
admittance. The fixture compensation includes OPEN and SHORT
compensation measurements. For more information on fixture compensation,
also refer to the Operation Manual of the Impedance Analyzer.
Generally, there is no need to perform load compensation. However, if you
have a suitable standard device and require extremely accurate
measurements, perform load compensation as needed.
The following procedure is for measurement of the compensation data with the
42941A.

Performing Fixture Compensation

Attach the probe adapter to be used for your measurement and perform fixture
compensation.
1. Press the [Cal] key to bring up the Calibration Menu.
2. Press the Fixture Compen key to bring up the Fixture Compensation Menu.
18
Keysight 42941A Impedance Probe Kit