Operation
Fixture Compensation
Performing Open Compensation
1. Attach the probe adapter to the probe but leave it disconnected from a
Figure 3-5
Performing Open Compensation
2. Press the OPEN key to start the OPEN compensation data measurement.
Keysight 42941A Impedance Probe Kit
device.
When the OPEN compensation data measurement is completed, the softkey
label Open [OFF] (if it is off) changes to Open [ON].
19