Operation

Impedance Analyzer Setup

3 Operation
This chapter describes the proper methods for setting up the Impedance
Analyzer, connecting the probe adapter, fixture compensation with the
42941A, and DUT measurement.
Impedance Analyzer Setup
Before beginning measurements, you should perform the adapter setup to
extend the calibration plane from the surface of the four-pair terminal to the tip
of the probe. Also refer to the Operation Manual of the Impedance Analyzer for
more information on the adapter setting.
For adapter setup, use the furnished short and load devices.
When handling the 42941A, care must be taken not to give it any
mechanical shock, which may cause damage to the fixture. Never give any
mechanical shock to the probe.

Adapter Setup

Connect the 42941A to the Impedance Analyzer as shown in
perform the adapter setup described below.
1. Leave the Impedance Analyzer for more than 30 minutes after turning it on
2. Press the [Cal] key to bring up the Calibration Menu.
3. Press the Adapter [ ] key to bring up the Adapter Setup Menu. [ ] should
4. Select PROBE 42941A. When selected, the softkey label will change to
5. Press the Adapter Setup key to bring up the Adapter Setup Menu.
6. Leave the 3.5-mm port of the 42941A open (no connection). There is no
7. Press the Phase [-] key to start the phase compensation data
Keysight 42941A Impedance Probe Kit
to allow the Impedance Analyzer to warm-up.
indicate current settings.
Adapter [42941A].
OPEN standard for the 42941A.
measurement. About 1 minute later, the phase compensation data
measurement is completed and the softkey label changes to Phase [Done].
Press [Save Phase].
Figure 3-1
and
13