Installing Robin® Mover; Inspection / Check Of The Rail System - Invacare EC-Track Installation & Technical Description

Rail system for ceiling hoist robin/robin mover
Table of Contents

Advertisement

Available languages
  • EN

Available languages

Invacare® EC-Track®
4. Locate the hoist trolley at the Ø 32 mm round recess
in the rail.
5. Lift the hoist and push the suspension pin through
the trolley opening.
6. Turn the hoist by 90 degrees to fix it in the trolley.
7. Connect the hand control.
8. Adjust the length of the emergency stop/lowering cord
to within the user's reach.
If there is excessive resistance, or if the trolley
unintentionally rolls by itself due to imbalance in
the rail suspensions, adjust the friction brake using
a 4 mm Allen key.
3.11.2 Installing Robin® Mover
The trolley is built-in on the hoist, and can therefore not
be pre installed in the rail.
Tools: 4 mm Allen key; 17 mm socket wrench
26
1. Remove the end stop.
2. Insert the wheels of the hoist into the rail.
The direction the hoist faces does not matter
during installation.
3. Install the end stop.
Tighten the bolt to 30 Nm.
4. If too much resistance is experienced when the
hoist is pulled, or the hoist moves too freely and
unintentionally, adjust the friction brake.
5. Connect the hand control.
6. Adjust the length of the emergency stop/lowering cord
to within the user's reach.
3.12 Inspection / check of the rail
system
CAUTION!
After the installation of a rail system, the entire
system must be inspected.
– The inspection of a rail system must only be
carried out by a suitably qualified person.
– The inspection must follow the Safety
inspection checklist / Rail system in the
Invacare® Robin® Service Manual.
1586749-C

Advertisement

Table of Contents
loading