5.2.2.1 Requirements ............... Assembly ........................ Scope of delivery ..................6.1.1 Scope of supply Sim Multi Bench with supply module "Sim Multi Basis" ..................... 6.1.2 Scope of supply Sim Multi Bench with supply module "Sim Multi Basis II" ..................Assembly of components................
Page 3
Dentsply Sirona Table of Contents Installation instructions Simulation Unit Sim Multi Bench 6.2.4 Assembly of dentist element ............6.2.5 Assembly of external device connections (water / air) (option)..6.2.6 Assembly of phantom head ............6.2.6.1 Assembly of track for phantom head ......
Page 4
Table of Contents Dentsply Sirona Installation instructions Simulation Unit Sim Multi Bench 7.7.3.4 Connecting the connection cable ......... 7.7.3.5 Adjusting the brakes ............. 7.7.3.6 Adjusting the pneumatic spring on the support arm ..7.7.3.7 Attaching panels ............Installing the monitor..................
Dentsply Sirona 1 Structure of the document Installation instructions Simulation Unit Sim Multi Bench 1.1 Identification of the danger levels Structure of the document Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions.
For reasons of product safety, this product may be operated only with original Dentsply Sirona parts or parts expressly approved by Dentsply Sirona. The user assumes the risk of unforeseen modifications and of using non-approved parts and accessories.
Dentsply Sirona 2 General safety information Installation instructions Simulation Unit Sim Multi Bench 2.4 Electrostatic discharge Electrostatic discharge Use an ESD wrist band. Connect it to the protective ground conductor. Please comply with the usual precautionary measures for handling printed circuit boards (ESD). Make sure you touch a ground point to discharge yourself prior to touching the components.
3 Preparations Dentsply Sirona 3.1 Required tools and measuring equipment Installation instructions Simulation Unit Sim Multi Bench Preparations Required tools and measuring equipment Tools Measuring equipment ● Fourfold measuring head for the exact display of operating pressures REF: 36 66 336 66 82 368 D3684 D3684.031.01.01.02 08.2019...
Page 9
Dentsply Sirona 3 Preparations Installation instructions Simulation Unit Sim Multi Bench 3.1 Required tools and measuring equipment Supplement for the LEDview Plus operating light ● Retaining ring pliers for outside rings (shaft) Supplement for the assembly of the support tube...
4 Delivery Dentsply Sirona 4.1 Checking the delivery Installation instructions Simulation Unit Sim Multi Bench Delivery Checking the delivery All simulation units are carefully checked and packed prior to shipment. Please carry out an incoming inspection of the equipment in order to make sure that it was not damaged during transport.
Dentsply Sirona 4 Delivery Installation instructions Simulation Unit Sim Multi Bench 4.2 Scope of delivery Scope of delivery ● Dentist element and supply module ● Accessories ● Support tube ● Phantom head ● Operating light - (A) Operating light - (B) Operating light support arm...
We recommend performing an inspection of the circumstances on location at least 4 weeks prior to installation. The checklist should help you when doing this. This can help ensure a smooth procedure on the day that the Sim Multi Bench simulation unit is actually in- stalled.
Dentsply Sirona 5 Installation requirements Installation instructions Simulation Unit Sim Multi Bench 5.2 On-site installation On-site installation 5.2.1 Substrate, floor Unevenness The floor must be level and horizontal in accordance with DIN 18 202. 5.2.1.1 Load-bearing capacity of the floor Load capacity The minimum load-bearing capacity of the floor must be 0.25 N/cm...
(0.08 mm) (0,1 mm) * = For use with an Air Venturi air jet pump ** = only Sim Intego, Sim Multi Bench, Sim Bench Water hardness 1,4 mmol l (= 8° dH ) 66 82 368 D3684 D3684.031.01.01.02 08.2019...
Page 15
Dentsply Sirona 5 Installation requirements Installation instructions Simulation Unit Sim Multi Bench 5.2 On-site installation 1. Check the distances of the installation field from the walls according 3/8" to the installation plan. max. 2. Check the cable lengths* of the mains cable or control cable for the max.
● Check the shipment for completeness based on the attached “scope of delivery” checklist. The furniture is supplied by an external supplier and does not form part of the scope of delivery. 6.1.1 Scope of supply Sim Multi Bench with supply module "Sim Multi Basis" Item No. Designation Dentist element...
Dentsply Sirona 6 Assembly Installation instructions Simulation Unit Sim Multi Bench 6.1 Scope of delivery 6.1.2 Scope of supply Sim Multi Bench with supply module "Sim Multi Basis II" Item No. Designation Dentist element Supply module with front cover Phantom head...
Installation instructions Simulation Unit Sim Multi Bench Assembly of components IMPORTANT The Sim Multi Bench components are installed in the designated items of furniture. This furniture is designed specifically for the project. As such, the dimensions may vary depending on the design.
Dentsply Sirona 6 Assembly Installation instructions Simulation Unit Sim Multi Bench 6.2 Assembly of components NOTE Risk of damaging an existing floor heating system Fastening holes may damage an existing floor heating system. ➢ Determine the location of the heating with the heating system switched on.
6 Assembly Dentsply Sirona 6.2 Assembly of components Installation instructions Simulation Unit Sim Multi Bench 6.2.2 Assembly of supply module "Sim Multi Basis II" Attach the supply module on the base box to the side of the furniture using 3 screws M6x12 (A) as shown in the image.
Dentsply Sirona 6 Assembly Installation instructions Simulation Unit Sim Multi Bench 6.2 Assembly of components Upon completion of installation, see chapter 7, secure the front cover (A) with 4 screws (see image). (Screws form part of the furniture) IMPORTANT Subsequent assembly is shown below using the example of the simulation unit with the supply module "Sim Multi Basis".
6 Assembly Dentsply Sirona 6.2 Assembly of components Installation instructions Simulation Unit Sim Multi Bench 6.2.4 Assembly of dentist element 1. Secure the guide track (A) using 8 screws M6x16 (C) in the designated screw holes from left to right as shown in the image.
Page 23
Dentsply Sirona 6 Assembly Installation instructions Simulation Unit Sim Multi Bench 6.2 Assembly of components 5. Assemble bump stops; see image IMPORTANT The position of the bump stops can vary slightly. 66 82 368 D3684 D3684.031.01.01.02 08.2019...
6 Assembly Dentsply Sirona 6.2 Assembly of components Installation instructions Simulation Unit Sim Multi Bench 6.2.5 Assembly of external device connections (water / air) (option) Insert the external device connection installation kit with cover into the designated recess on the table top and secure from below with 4 nuts and washers (see image).
Dentsply Sirona 6 Assembly Installation instructions Simulation Unit Sim Multi Bench 6.2 Assembly of components 6.2.6 Assembly of phantom head 6.2.6.1 Assembly of track for phantom head Mount the track on the side of the furniture as shown in the image. The size and number of screws provided (M6 or M8) depend on the specific version.
6 Assembly Dentsply Sirona 6.2 Assembly of components Installation instructions Simulation Unit Sim Multi Bench 6.2.6.2 Assembly of phantom head on rod Insert the torso into the track mount Secure the torso using a clamping lever and fix with a setscrew (A).
Assembly of light and monitor, see chapter 7.7 and chapter 7.8 Necessary furniture dimensions 1200 mm x 600 mm x 840 mm for installation of the Sim Multi Bench components. An opening of 700 mm x 140 mm in the floor plate is to be created for the on-site connector panel.
7 Installation Dentsply Sirona 7.1 Installation of media supply Installation instructions Simulation Unit Sim Multi Bench Installation Installation of media supply IMPORTANT Before commencing the subsequent installation work, the rear panel, or, depending on the furniture design, the front cover of the "Sim Multi Basis II"...
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.1 Installation of media supply Compressed air line ü Compressed air must be available (compressor switched on). ➢ Open the compressed air corner valve for 30 seconds. >30 sec Water path ü...
Page 30
7 Installation Dentsply Sirona 7.1 Installation of media supply Installation instructions Simulation Unit Sim Multi Bench Connecting the air and water supply ü The positions of the corner valves correspond to Sirona's specifications. ü The air and water paths have been cleaned of chips and filings. See section "Cleaning compressed air and water lines“.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.2 Connecting drain and suction hoses Connecting drain and suction hoses NOTE It is possible to confuse the suction pipe with the drain pipe. This could cause waste water to be sucked into the suction machine.
7 Installation Dentsply Sirona 7.2 Connecting drain and suction hoses Installation instructions Simulation Unit Sim Multi Bench 7.2.2 Connecting the suction hose 1. Grease the O-ring with grease (Unisilkon LCA 3801). 2. Insert the end piece of the suction hose (B) (large hose diameter).
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.3 Electrical connections Electrical connections 7.3.1 Making the electrical connection for the building installation 7.3.1.1 Service/Installation personnel qualifications Installation/Startup work must be carried out only by electricians or by trained and qualified personnel or by personnel authorized by Dentsply Sirona.
7 Installation Dentsply Sirona 7.3 Electrical connections Installation instructions Simulation Unit Sim Multi Bench 7.3.2 Connecting the power cord The unit's interface for building installation is formed by connector A and socket B. IMPORTANT The operator is responsible for connecting the on-site power cable to the socket B.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.3 Electrical connections 3. Plug in socket B into connector A. 4. Secure socket B with screw C (M3x16) 5. Routing power cable D. Secure the power cable to the designated cable clamp with a cable tie (cable relief) and lay the cable inside the supply module.
7 Installation Dentsply Sirona 7.3 Electrical connections Installation instructions Simulation Unit Sim Multi Bench 7.3.3 Connecting the control cable for the suction machine Connection of the cable for suction machine connection A to terminal X4 of the DWB or DWE board.
Page 37
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.3 Electrical connections 2. Assemble cable connection C on one of the cable entries in the junction box. IMPORTANT Depending on the cable installation of the on-site control cables, a suitable cable entry in the junction box can be selected to assemble the cable connection.
7 Installation Dentsply Sirona 7.4 Installation of dentist element Installation instructions Simulation Unit Sim Multi Bench Installation of dentist element 7.4.1 Hydropneumatic connections Routing air hoses and water hoses Supply module "Sim Multi Basis" 1. Route the water hose W12 and the air hoses A11, A12, A13 and A15 in the supply module.
Page 39
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.4 Installation of dentist element Connections in the supply module 1. Attach the water hose W12 to the distribution box VW1. 2. Attach the air hose A15 to the distribution box VLb / VLa.
Page 40
7 Installation Dentsply Sirona 7.4 Installation of dentist element Installation instructions Simulation Unit Sim Multi Bench Connection for hose set for dentist element / Adapter for pneumatic foot control 1. Attach the hoses of the hose set for dentist element (A) to the adapter for pneumatic foot control (B) as follows: –...
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.4 Installation of dentist element Option: with electrical foot control Connections to valve block Valve block Hose set for dentist element with safety plate 1. Insert the air hoses A11, A12, and A13 into the valve block (A).
7 Installation Dentsply Sirona 7.4 Installation of dentist element Installation instructions Simulation Unit Sim Multi Bench 7.4.2 Electrical connections 7.4.2.1 Connections from the dentist element to the supply module "Sim Multi Basis" Connecting cable L546 1. Route cable L546 to the DWB/DWE board in the supply module.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.4 Installation of dentist element 7.4.2.2 Connections from the dentist element to the supply module "Sim Multi Basis II" Connecting cable L546 1. Route cable L546 to the DWB board in the supply module.
7 Installation Dentsply Sirona 7.4 Installation of dentist element Installation instructions Simulation Unit Sim Multi Bench 7.4.3 Connecting the electrical foot control 7.4.3.1 Supply module "Sim Basis" 1. Route the cable L548 and the protective ground wire of the foot control in the supply module.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.4 Installation of dentist element 7.4.3.2 Supply module "Sim Basis II" 1. Route the cable L548 and the protective ground wire of the foot control in the supply module.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.6 Installation of suction surface suction hoses Installation of suction surface suction hoses Attaching the suction hoses 1.Grease the O-rings (A). Instrument holder for suction handpieces Position 1 Position 2...
Installation instructions Simulation Unit Sim Multi Bench Installation of the operating light 7.7.1 Assembly position of lamp post Assembly position of the operating light with Sim Multi Bench with patient simulator with mechanical height adjustment, horizontally adjustable dentist element and LEDview Plus IMPORTANT Alternatively, the light crown LEDlight Plus can be assembled.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7.7.2 Assembling the support sleeve M6x60 ➢ Secure the support sleeve (B) with 4 screws M6x60 (included in the scope of delivery) to the retaining plate (I), through the designated drill holes on the table top and tighten the nuts.
7 Installation Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench 7.7.3 Assembling the operating light LEDview Plus/LEDlight Plus CAUTION Risk of crushing when assembling the light. The support arm structure is designed to swivel.
Page 51
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 4. Remove the pin (C) and push the cover (E) over the support tube (F) beyond the threaded hole of the pin (C). Secure the cover in place (E) with adhesive tape and screw the pin (C) back into the threaded hole.
Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench 10. Align the support tube with a water level by adjusting the 4 setscrews (A). To do this use a right-angled 6 mm Allen key.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7.7.3.2 Attaching the lighting support arm to the support column Installation material ● Installation material support arm operating light Installation of rigid support arm 1.
Page 54
7 Installation Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench Installation of flexible support arm 1. From below, feed the joint of the flexible support arm into the vertical axis of the rigid support arm.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7.7.3.3 Assembling light on light support arm 1. Insert the brake (E) into the joint from above. IMPORTANT! The coating (D) of the brake (E) must face downward during installation.
Page 56
7 Installation Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench 4. Engage brake (B) on the support arm. 5. Insert cable L573 through the front support arm joint and the joint of the light.
Page 57
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7. Install the cover panels of the handle axis holder on the light. Attach handle sleeve (LEDview Plus) 1. Slowly pull the handle sleeve over the handle adapter until it audibly clicks into place on the handle adapter.
7 Installation Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench 7.7.3.4 Connecting the connection cable Securing cable L573 onto the fixed support arm 1. Pull cable L573 through the joint from below. 2. For the next step use a climbing aid.
Page 59
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7.7.3.4.1 Connecting the connecting cable of the operating light in the supply module "Sim Module Basis" Connecting the light LEDview Plus 1. Route connecting cable L573 through the support tube to the supply module.
Page 60
7 Installation Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench Connecting the light LEDlight Plus 1. Route connecting cable L573 through the support tube into the supply module. L573 2. Connect plug X45 of cable L578 supplied to connecting cable L573 of the light with an overhung connection.
Page 61
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7.7.3.4.2 Connecting the connecting cable of the operating light in the supply module "Sim Module Basis II" Connecting the light LEDview Plus 1. Route connecting cable L573 through the support tube to the supply module.
Page 62
7 Installation Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench Connecting the light LEDlight Plus 1. Route connecting cable L573 through the support tube into the supply module. 2. Connect plug X45 of cable L578 supplied to connecting cable L573 of the light with an overhung connection.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7.7.3.5 Adjusting the brakes Adjusting the brake on the support column IMPORTANT This brake is not preset. Adjust it to the required brake force.
7 Installation Dentsply Sirona 7.7 Installation of the operating light Installation instructions Simulation Unit Sim Multi Bench Adjusting the brake between the support arm and lighting crown 1. Increasing brake force at position (A): Using an Allen key turn the screw (H) inwards.
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.7 Installation of the operating light 7.7.3.7 Attaching panels 1. Attach the panels (A) from both sides. 2. Attach the panels (B). Ensure that the guide pins (F) lock into the holes (G).
7 Installation Dentsply Sirona 7.8 Installing the monitor Installation instructions Simulation Unit Sim Multi Bench Installing the monitor 7.8.1 Optional: Installation without operating light Declaration If a monitor without an installed operating light is to be operated on the simulation unit, a support tube must first be installed.
Page 67
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.8 Installing the monitor Fastening the monitor 1. Fix the monitor to the holders (C) on the adapter plate using the screws (B). 2. Screw the monitor down to the outside positions (D) with the screws (B) and with additional screws.
Page 68
7 Installation Dentsply Sirona 7.8 Installing the monitor Installation instructions Simulation Unit Sim Multi Bench Connecting cables 1. Insert IEC socket (A) of cable L529 into the IEC connector. 2. Plug cable L569 (HDMI cable) in socket (B). Attaching the retaining plate 1.
Page 69
Dentsply Sirona 7 Installation Installation instructions Simulation Unit Sim Multi Bench 7.8 Installing the monitor 2. Set the brake force for the angle of the monitor using screws (B). Attaching panels 1. Secure both covers (A) and (B) to the adapter plate using one screw (C) each.
7 Installation Dentsply Sirona 7.8 Installing the monitor Installation instructions Simulation Unit Sim Multi Bench 7.8.3.2 Connections to the supply module "Sim Multi Basis II" ü The main switch is in a voltage-free state. 1. Route cable L529 (A) to the main switch (B) in the supply module.
Dentsply Sirona 8 Initial startup Installation instructions Simulation Unit Sim Multi Bench 8.1 Connecting the instrument hoses to the dentist element Initial startup Connecting the instrument hoses to the dentist element 8.1.1 Dentist element Connect the instrument hoses (according to order) onto the dentist element from left to right.
SiroCam AF (not for USA/ Canada) The Sim Multi Bench simulation unit can be fitted with a maximum of two motors. A maximum of one scaler and one intraoral camera can be connected. Changes in the instrument positions can be made only by your service technician.
Dentsply Sirona 8 Initial startup Installation instructions Simulation Unit Sim Multi Bench 8.2 Starting up the simulation unit Starting up the simulation unit 8.2.1 General Important notice: changing a 3-way syringe hose NOTE Uncontrolled water and air leaks When the simulation unit is switched on, uncontrolled air and water leaks can occur.
8 Initial startup Dentsply Sirona 8.3 Checking and adjusting the operating pressures of the instruments Installation instructions Simulation Unit Sim Multi Bench Checking and adjusting the operating pressures of the instruments 8.3.1 Air/water factory settings IMPORTANT Factory setting The driving air pressure can be set exactly using the four-way measuring head (Order No.
Dentsply Sirona 8 Initial startup Installation instructions Simulation Unit Sim Multi Bench 8.3 Checking and adjusting the operating pressures of the instruments 8.3.2 Setting the electric motors The cooling air pressure for the electric motor results from the driving air setting of the turbine.
Page 78
8 Initial startup Dentsply Sirona 8.3 Checking and adjusting the operating pressures of the instruments Installation instructions Simulation Unit Sim Multi Bench 8.3.3.1.1 Opening the dentist element 1. Unscrew the 3 screws from the side of the cover flap of the dentist element.
Page 79
Dentsply Sirona 8 Initial startup Installation instructions Simulation Unit Sim Multi Bench 8.3 Checking and adjusting the operating pressures of the instruments 4. Lift up the intermediate floor to access the module. 66 82 368 D3684 D3684.031.01.01.02 08.2019...
8 Initial startup Dentsply Sirona 8.3 Checking and adjusting the operating pressures of the instruments Installation instructions Simulation Unit Sim Multi Bench 8.3.3.2 Adjusting the driving air To do this, use a four-way measuring head. Driving air Spray air (non-adjustable; only used for checks).
Dentsply Sirona 9 Dismantling and reinstallation Installation instructions Simulation Unit Sim Multi Bench 9.1 Service personnel qualifications Dismantling and reinstallation When dismantling and reassembling the unit, proceed according to the installation instructions for new installation in order to guarantee its functioning and stability.
Need help?
Do you have a question about the Sim Multi Bench and is the answer not in the manual?
Questions and answers