Specifications; Cleaning And Disinfecting Stimpod - Xavant STIMPOD NMS 410 Manual

Quantitative nmt monitor, precision nerve locator
Table of Contents

Advertisement

5.2) Specifications

Operating Modes:
Current Range
Pulse Width Options
Maximum Stimulation Voltage
Stimulus
Stimulating Frequency
Load Impedance
Technical Specifications
Device Classification
Power Supply
Power Consumption
Waveform
Weight
Dimensions
Operating Temperature
Storage and Transport Temperature
Operating Humidity
Transport and Storage Humidity
Operating Atmospheric Pressure
Transport and Storage Atmospheric Pressure
5.3) Cleaning and Disinfecting STIMPOD NMS 410/450
Cleaning:
Soap and water, applied with a damp cloth is suitable to clean and disinfect the STIMPOD. It is imperative that no moisture penetrates the STIMPOD.
Disinfecting:
Any commercially available methanol - free disinfectant in an ethyl alcohol base can be used for disinfection.
34
Nerve Locating Mode
NMS 410/450X
0.00 - 5.00 mA ± 5%
0.05ms, 0.1ms, 0.3ms, 0.5ms, 1ms ± 5%
100V
Monophasic square wave
1Hz , 2Hz, 5Hz ± 5%
0 kOhm - 20 kOhm
NMS 410/450X
Class IIa, Type BF
4 x AAA alkaline batteries
17mA
Constant Current, Monophasic Square
Wave
130g
145mm x 90mm x 30mm
10 - 40 ° Celsius
0 - 50 ° Celsius
90% Relative Humidity
90% Relative Humidity
50 – 106kPa
50 – 106kPa
Nerve Mapping Mode
NMS 410/450X
0 - 20mA ± 5%
0.05ms, 0.1ms, 0.3ms, 0.5ms, 1ms ± 5%
400V
Monophasic square wave
1Hz , 2Hz, 5Hz ± 5%
0 kOhm - 20 kOhm
NMT Mode
NMS 450X
0 - 80mA ± 5%
0.2ms ± 5%
400V
Monophasic square wave
1Hz, 2Hz, 5Hz, 50Hz, 100Hz ± 5%
0 kOhm - 5 kOhm