Current Setting
Adjust using wheel
Stimulation Mode

3.1) Adjusting the Current

Default Current range: 0 - 80 mA adjustable in 5mA increments.
Circle the Wheel to adjust current.
When the Stimpod starts up in NMT mode, it will revert to the NMT current that was used
previously. In order to adjust the current, circle the adjustment wheel, the current setting
will flash and start to adjust accordingly. At the same time a warning screen will appear
prompting you to press ' e nter' to confirm. Press ' e nter' to confirm the current change
within 2 seconds.

3.2) Adjusting Stimulation Mode

Stimulation mode TOF, DB, PTC, TET, TWI, SMC and Auto
Default: TOF
Press Menu/Mode button to toggle between different Stimulation Modes.
1
2
1
2
3.3) Adjusting Twitch / Tetanus Frequency
Twitch Mode: Options (Stimulating Frequency): 1Hz, 2Hz, 5Hz
NMS 450X
NMS 450X
Default: 2 Hz
Press Enter/Hz button to toggle between different stimulation frequencies.
Tetanus Mode: Options: 50Hz, 100Hz
Default: 50Hz
Press Enter/Hz button to toggle between different stimulating frequencies.
Warning:
Notifies the user of a discrepancy
3
between current setting and average current
of actual stimulus.
If sound is enabled:
STIMPOD will beep
4
every time that a stimulus was delivered. The
pitch of the sound will follow the current
intensity level.
Average current of actual stimulus
5
21