The Stimpod is designed to be compatible with a standard ECG electrode, however,
for high currents the use of a dedicated NMT electrode such as the Xavant XT45008
is recommended.
Electrodes that have current densities exceeding 2mA/cm2 may require special
attention of the operator.
This product must be stored at 0 – 50°C.
This product must be transported in the carry case provided.
This product and all the accessories have been certified latex free.
Inspect all parts for any damage or manipulation. Never use any damaged or
manipulated part!
If an electrically conductive surface of the Stimpod device or its cables are exposed,
such electrically conductive surface may shock a person handling it. Do not use such
a device or accessory, please contact the manufacturer for repair.
The refractory period delay is set at a default value to prevent the user from
repeating stimulation while the nerve synapse is recovering from effects of the
previous stimulation. A refractory period of less than 12 seconds in TOF mode is not
advisable as measurements might not represent the effect of blocking agents on the
neuromuscular junction.
Application Specification:
The patient population includes patients of all ages, weight and nationality. Patient
health and state is described in contraindications, warnings and cautions.
The user must be a medical professional with knowledge of anatomy.
The use environment requirements of the device such as the medical practice and
operating room is described in the guidance and manufacturers declaration.
The device can be used on any part of the body except for limitation described in
warnings and cautions or identified in section 3 for the NMT mode.
Warranty:
The Stimpod (device only) carries a 24 Month Warranty against defects, provided that
the device was used in accordance with the operating instructions.
The cables included in the Stimpod Kit carry a 6 Month Warranty against defects,
provided that they were used in accordance with the operating instructions.
The Stimpod enclosure should not be opened under any circumstances. Opening
the unit will void the warranty.
STIMPOD (NMS 410/450X) conforms to the following
standards:
IEC 60601-1, IEC 60601-2-10
IEC 60601-1-2: CISPR 11 Group1 class A; IEC 61000-4-2; IEC 61000-4-3
ISO 13485, Directive 93-42-EEC