  • Page 1 PRODUCT CODE: XT-46006-EN XM400-31A04-07...
  • Page 3 • Do not use the STIMPOD NMS460 in close proximity to equipment that produces Xavant Technology Pty (Ltd) strong electromagnetic fields, such as high frequency surgical equipment. The Unit 102, The Tannery Industrial Park, 309 Derdepoort Rd...
  • Page 4 • Stimulation should not be applied across or through the head, directly on the • Since the effects of stimulation of the brain are unknown, stimulation should not eyes, covering the mouth, on the front of the neck, (especially the carotid sinus), be applied across the head, and electrodes should not be placed on opposite sides or from electrodes placed on the chest and the upper back or crossing over the of the head.
  • Page 5 Guidance and manufacturers declaration – electromagnetic emissions– for all equipment and systems The STIMPOD NMS460 is intended for use in electromagnetic environment specified below. The customer or user of the STIMPOD NMS460 should ensure that it is used in such an environ- ment Emission Test Compliance...
  • Page 6 Guidance and manufacturer’s declaration – electromagnetic immunity The STIMPOD NMS460 is intended for use in the electromagnetic environment specified below. The customer or the user of the STIMPOD NMS460 should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment –...
  • Page 7 Recommended separation distances between portable and mobile RF communications equipment and the STIMPOD NMS460 The STIMPOD NMS460 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the STIMPOD NMS460 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the STIMPOD NMS460 as recommended below, according to the maximum output power of the communications equipment.
  • Page 8: Table Of Contents

    Contents Page Getting to know the STIMPOD NMS460 1.1) Device Description 1.2) Accessories 1.3) Device Layout 1.4) Screen Layout 1.5) Adjusting Settings 1.6) Open Circuit Detection 1.7) Countdown Timer 1.8) Shutdown 1.9) Symbols on device battery clip Setting up Device Defaults 2.1) Languages 2.2)
  • Page 9: Getting To Know The Stimpod Nms460

    The second waveform is a Hybrid RF waveform which consists of a Monophasic Square Wave with a superimposed Radio Frequency waveform. This waveform is proprietary and is unique to the STIMPOD NMS460. For more information on the benefits of using this modality please visit www.xavant.com.
  • Page 10: Device Layout

    1.3) Device Layout Cable Connector • Insert the Cable Multi Functional Clip Enter / Frequency Button Display • Press to toggle between Stimulating Frequencies. • Press to Enter in setup menu. Stimulating LED indicator • Flashing Green: Stimulus delivered. • Flashing Red: Open Circuit detected.
  • Page 11: Screen Layout

    1.4) Screen Layout • Indicators • Countdown timer Square Wave Hybrid RF • Current Setting Speaker Volume Adjust using Wheel. • Pulse Width Setting Battery Status Adjust using Menu / Pulse Width button. • Stimulating Frequency Setting • Warning Screen Adjust using Enter / Hz button.
  • Page 12: Adjusting Settings

    1.5) Adjusting Settings 1.7) Countdown Timer • Start the stimulation by pressing the play/pause button. Adjusting the Current • The device will automatically start stimulation at the set Current Range: 0 – 30 mA adjustable in 1mA increments frequency and the countdown timer will start. Circle the Wheel to adjust the current •...
  • Page 13: Setting Up Device Defaults

    Setting up Device Defaults Access the Setup Menu by pressing Hybrid RF Mode Menu/Pulse Width and holding the The “ “ indicator in the main window tells the user that the device is in Button. The setup menu allows the Hybrid RF Mode.
  • Page 14: Pulse Width Options

    2.3) Pulse Width Options 2.5) Countdown Timer DEFAULT: 0.1ms DEFAULT: 10 minutes Options: 0.1ms, 0.2ms Options: Adjustable from 00:01 to 59.59 Use this setup function to activate the preferred pulse widths. Use the Wheel to navigate to the COUNTDOWN TIMER .
  • Page 15: Backlight

    2.7) Backlight DEFAULT: 5 Seconds Options: OFF, 5 seconds, 60 seconds, and Always On Use the Wheel to navigate to BACKLIGHT in the main menu. Press enter to select. BACKLIGHT Within the Menu, the radio button next to the selected setting will show that it has been activated. Scroll down the menu to EXIT MENU and press enter.
  • Page 16: Technical Notes

    Technical Notes 3.1) Performance Test Short-circuit the Nerve Mapping probe and the ECG connector. • The following screen should appear on the display. Before operating and using the device a performance test must be carried out at the site of use. The performance test described below is in compliance with the German §...
  • Page 17: Specifications

    3.2) Specifications Current Range: 0 - 30mA ± 5% Pulse Width Options: 0.1ms, 0.2ms Maximum Stimulation Voltage: 220V Waveform: Monophasic square wave/ HYBRID RF wave Stimulating Frequency: 1Hz , 2Hz, 5Hz, 10Hz ± 5% Load Impedance: 0 kOhm - 5 kOhm Device Classification: Class IIa, Type BF Power Supply:...
  • Page 18: Products And Accessories

    Products & Accessories Product Product Code NMS 460 PERIPHERAL NERVE STIMULATOR ONLY XT-46000 (No cables or accessories) NMS 460 PERIPHERAL NERVE STIMULATOR KIT XT-46001 (Including Mapping Probe, Carry Case, IFU) NERVE MAPPING CABLE XT-46003 CARRY CASE XT-41002 POLE/ WALL MOUNT XT-41007...
  • Page 20 Unit 102, The Tannery Industrial Park, 309 Derdepoort Rd Silverton, Pretoria, South Africa, 0184 Tel: +27 (0) 12 743 5959, Fax: +27 (0) 86 547 0026 E-mail: support@xavant.com, Web: www.xavant.com...