A) Adjusting The Current In Loc Mode - Xavant STIMPOD NMS 410 Manual

Quantitative nmt monitor, precision nerve locator
Table of Contents

Advertisement

When using the Nerve Locating / Mapping Cable
Current Setting
Adjust using wheel
Pulse Width Setting
Adjust using Menu / Pulse width button

2.1a) Adjusting the Current in LOC mode

Current Mode Options: Linear Mode, Non-Linear Mode
Default: Linear
Linear Mode:
The Linear Mode is called linear because one ' c lick' on the Wheel will correspond with one
increment as set in the specific current range. The Linear Mode allows the user to select
individual incrementing options for the three different current ranges.
Default Current range:
0.00 - 5.00mA adjustable in the following default increments:
0.0 - 0.6mA Default 0.1mA
16
1
2
1
0.6 - 2.0mA Default 0.2mA
2.0 - 5.0mA Default 0.5mA
Increments can be adjusted in the Setup Menus
Circle the Wheel to adjust current.
Non-Linear Mode:
NMS 450X
NMS 450X
The Non-Linear Mode facilitates the non-linear nature of the current intensity versus the
distance from the nerve. This mode allows the user to define 20 adjustment positions in
terms of Current (mA) and Pulse Width (ms). If correctly implemented each adjustment
position should afford the user with a relatively linear progression in terms of the distance
from needle tip to the nerve.
Warning:
Notifies the user of a discrepancy
3
between current setting and average current
of actual stimulus.
If sound is enabled:
STIMPOD will beep
4
every time that a stimulus was delivered. The
pitch of the sound will follow the current
intensity level.
Average current of actual stimulus
5
Proximity Indicator
6
Stimulating Frequency Setting
7
Adjust using Enter / Hz width button