Xavant STIMPOD NMS 410 Manual page 33

Quantitative nmt monitor, precision nerve locator
Table of Contents

Advertisement

5.1.3) Neuromuscular Transmission Monitoring Mode
(NMT Mode) (NMS450X only)
Insert the NMBA Cable.
The following screen should appear on the display.
Offener Stromkreis
!
Ensure that the device is in 'TOF' mode.
Short-circuit the red and black electrode connectors.
Use the adjusting wheel and increase the current to 80mA.
Press the 'play/pause' button while shaking the accelerometer.
The NMS450 should respond as follows:
The LED should flash GREEN in accordance with the four stimulations.
Each stimulation should be accompanied by an audible 'beep' .
In 'Diagnostic window' four bars of different heights should indicate that the
accelerometer detected movement.
Monitor the actual current delivered to ensure that the warning sign does not appear.
Open Circuit
erkannt
Detected
NMS 450X
NMS 450X
NMS 450X
NMS 450X
Separate the red and black electrode connectors to cause an open circuit between
them.
Offener Stromkreis
!
Press the 'play/pause' button.
The following screen should appear on the display.
No audible feedback should be heard.
If the STIMPOD malfunctions in any one of these performance tests, it should
be checked by the relevant technical department in accordance with the test
instructions in the Technical Service Manual.
Equipment may only be repaired by the manufacturer or by an organisation expressly
authorised by the manufacturer.
Equipment does not require regular calibration.
Open Circuit
erkannt
NMS 450X
NMS 450X
Detected
NMS 450X
NMS 450X
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Stimpod nms 450x