EMI Warning
EMI Warning:
Warning shown when a high level of electromagnetic interference is detected.
Electrode Placement
Incorrect/Check electrode placement:
Warning shown when SMC is unable to find a supramaximal current value.
No Accelerometer
Data Received
No Accelerometer data received:
Warning shown when the accelerometer didn't respond.
Stimulation aborted
Stimulation aborted:
This warning is shown when:
SMC:
measured current is more than 10% less or more than the set current.
NMT:
Stimulation is interrupted by user input.
Incorrect
Component Error:
The STIMPOD has detected a component failure. Please send the device back to the
manufacturer for repair!