Technical Notes; Performance Testing - Xavant STIMPOD NMS 410 Manual

Quantitative nmt monitor, precision nerve locator
Table of Contents

Advertisement

5.1) Performance Test
Before operating and using the device a performance test must be carried out at the
site of use. The performance test described below is in compliance with the German § 5
MPBetreibV directive.
Insert the batteries and switch on the device.
The following screen should appear on the display.
NMS4X0
User Information
followed by
VX.00
NMS 450X
NMS 450X

Technical Notes

.
_ _ _
Insert Cable
Insert Cable
!
5.1.1) Nerve Locating Mode
Insert the Nerve Mapping/Nerve Locating Cable.
The following screen should appear on the display.
0.00
Open Circuit
!
Detected
5
Insert Cable
NMS 450X
NMS 450X
Open Circuit
Detected
31