Xavant STIMPOD NMS 410 Manual page 30

Quantitative nmt monitor, precision nerve locator
Table of Contents

Advertisement

Current
Position
(mA)
1
0.3
2
0.43
3
0.58
4
0.76
5
0.97
6
1.2
7
1.4
8
1.7
9
2
10
2.3
11
2.7
12
3
13
3.4
14
3.8
15
4.3
16
4.8
17
1.8
18
2.1
19
2.4
20
2.7
Table 1 Default non-linear settings
30
Pulse Width
Charge
(ms)
(µC)
0.1
0.03
0.1
0.043
0.1
0.058
0.1
0.076
0.1
0.097
0.1
0.12
0.1
0.14
0.1
0.17
0.1
0.2
0.1
0.23
0.1
0.27
0.1
0.3
0.1
0.34
0.1
0.38
0.1
0.43
0.1
0.48
0.3
0.54
0.3
0.63
0.3
0.72
0.3
0.81
These values can be viewed or changed in the Non-linear Settings menu. The charge value
for each or the twenty positions are displayed graphically, and the current and pulse width
value of each position can be viewed by scrolling to that position using the wheel.
To change the charge at a specific position, a 2 step process can be followed. Select the
position which should be changed using the wheel and press the enter button to start
step 1. The current is changed using the wheel in this step and the menu button will
cancel the change. When satisfied with the current, the enter button can be used to
continue to step 2: changing the pulse width. The pulse width for the selected position
can be changed using the wheel. The menu button will return to step 1 and the enter
button will accept the change and conclude editing.
step 1
step 2
NMS 450X
NMS 450X
NMS 450X
NMS 450X