SECTION 3 — BEAD RUPTOR SAMPLE TUBES
SECTION 3 — BEAD RUPTOR SAMPLE TUBES
3.2 ORDERING INFORMATION
50 mL PRE-FILLED BEAD MILL TUBES
BEAD MEDIA
TUBE SIZE
Ceramic
50 mL
TUBE SIZE
50 mL
50 mL Bead Tubes are not certified RNase/DNase Free, Maximum recommended speed is 5.0 m/s.

3.4 IMPORTANT SAMPLE TUBE INFORMATION

Dry, fresh or frozen samples can be homogenized with or without buffer using the Bead Ruptor 24
and Bead Ruptor sample tubes.
The intratube forces generated by the Bead Ruptor 24 during the homogenization process are up to 400 x g.
Intratube forces are dependent on the speed setting and the bead media present inside the tube. Omni pro-
vides unique tubes that are designed to withstand such high forces. It is not recommended to use tubes pro-
vided by other suppliers as these may not withstand to forces generated by the Bead Ruptor 24 and could
result in tube failure and loss of sample. Please reference the table below for Omni tube speed thresholds.
TUBE SIZE
SPEED
0.5ml
8 m/s
1.5ml
5 m/s
CAUTION:
2ml non reinforced tubes are not compatible with beads >2mm in diameter.
In addition to the intratube forces generated by the Bead Ruptor 24, sample heating can occur when process-
ing for multiple cycles or extended time periods. Excessive sample heating can result in tube leakage or failure.
Because the Bead Ruptor has the ability to run at up to 8 m/s, tube leakage and breakage could occur due
to heat and pressure build-up inside the tube.
In the event of tube leakage or tube breakage, Omni recommends the following solutions:
• Use the tubes as single-use or disposable
• Add more buffer to decrease the force applied by the beads to the tube walls
• Open the Bead Ruptor's lid for 5 to 10 minutes in order to cool down the tubes and tube carriage between cycles.
• Place the sample tubes in an ice bucket for few minutes in between cycles
prior to run additional cycles.
• Ensure that the fingerplate is tightened securely and the lock is engaged.
• Ensure tube caps are tightened completely
CAUTION:
DO NOT freeze Bead Ruptor Tubes in liquid nitrogen.
16
BEAD SIZE
2.8 mm
Hard Tissue Homogenizing Mix
Bulk 50 mL Screw Cap Sample Tubes (No Beads)
Empty sample tubes with caps. Beads are not included.
DESCRIPTION
Polypropylene conical screw cap
tubes with caps
TUBE SIZE
SPEED
2ml non reinforced
8 m/s
2ml reinforced
8 m/s
DESCRIPTION
QUANTITY
100
TUBE SIZE
SPEED
7ml
6.0 m/s
15ml
7.0 m/s
QUANTITY
ORDER NUMBER
50
19-6508
ORDER NUMBER
19-6650
TUBE SIZE
SPEED
30ml
5.0 m/s
50ml
5.0 m/s