SECTION 5 — PROGRAMMING

5.2 ADJUSTING PARAMETERS

ADJUSTING THE NUMBER OF CYCLES
Use the  and  arrow keys to highlight the "C"- Cycles option.
The user can change the Number of Cycles using the  and  Arrow buttons. The number of cycles ranges
from 1 to 10.
When the desired value is reached, the user can go to the next desired variable or go back up to Program to
begin the run.
ADJUSTING THE PAUSE OR "DWELL" BETWEEN CYCLES
Use the  and  arrow keys to highlight the "D"- Pause Dwell option.
The user can change the Dwell time between 2 or more cycles using the  and  Arrow buttons. The dwell
time between cycles ranges from 0 seconds to 9 minutes and 59 seconds in increments of 1 second.
When the desired value is reached, the user can go to Program to begin the run.
The unit will beep twice when the run is complete.
CAUTION:
Wait until the unit stops and displays "DONE—REMOVE SAMPLES" before opening the lid. The
drive hub can become very hot during operation. Use caution when removing the tube carriage.
24
20 o C
PROG 04
S =4.50
C=02
T=0:10
D=0:10
NUMBER OF CYCLES
20 o C
PROG 04
S =4.50
C=02
T=0:10
D=0:10
PAUSE DWELL