Symbols ..........................2 1.1.3 Target group ........................2 Setup methods according to DIN EN ISO 17664 ..................3 References ............................3 General instructions for the reprocessing of KaVo treatment units ..........4 2.2.1 Reprocessing overview Primus 1058 S/TM/C ..............5 ®...
Care instructions KaVo Primus® 1058 S/TM/C 1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide Requirement Read these instructions prior to first use to avoid misuse and prevent damage. Note Comply with the Instructions for Use of the Primus 1058 S/TM/C treatment unit! ®...
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.1 References 2 Setup methods according to DIN EN ISO 17664 2.1 References The reprocessing consists of the following basic steps: ▪ Cleaning and disinfection ( manually or automatically) ▪...
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units 2.2 General instructions for the reprocessing of KaVo treatment units Please comply with the general reprocessing instructions at all times. Detailed product-specific reprocessing instructions are provided subsequently.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units Manual drying ▶ Supply dry, clean and uncontaminated compressed air according to ISO 7494-2 only.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.3 Reprocessing of the surfaces of treatment unit / upholstery Reprocessing items/medical devi‐ Cleaning Disinfection Sterilisation Connection parts of the suction hose Patient element...
Page 9
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.3 Reprocessing of the surfaces of treatment unit / upholstery Approved disinfectants: ▪ Microcide AF liquid (Schülke & Mayr) ▪ FD 322 (Dürr) ▪ Incidin Liquid (Ecolab) ▶...
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit 2.4 Preparation of the dentist unit 2.4.1 Reprocessing of holder pad/anti-slip mat Cleaning, disinfection, and sterilisation of holder pad/anti-slip mat...
Page 11
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit CAUTION Damage caused by thermodisinfection. ▶ Do not thermodisinfect the handle of the TM table / cart. ▶ Clean the handle with a soft cloth and water, and wipe-disinfect it with an ap‐...
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Clean the handle under flowing water. ▶ Disinfect the white and grey handle by wiping or thermodisinfection (minor dis‐...
Note Wet the O-rings of the sleeves after sterilisation using silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replac‐ ing the cannula, the air channel must be blown out before the start of treatment to ensure that the air exiting during treatment is dry.
Page 14
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit Accessories required: ▪ Tap water 30°C ± 5°C ▪ Brush, e.g. medium-hard toothbrush ▶ Disassemble the gripping sleeve and cannula.
Page 15
Requirement The multifunctional handpiece has been prepared. ▶ Apply KaVo silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510) to the O-rings of the joint between the gripping sleeve and cannula. Use cotton buds for this purpose.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Hold the handpiece at the taper sleeve and take off the cannula with a slight twisting motion.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit 2.4.6 Servicing the turbine return air filter Cleaning the turbine return air filter Note The liquid collector must be checked weekly.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant unit Cleaning the kit for physiological saline solution When using saline solution, all lines conducting the solution must be rinsed direct‐...
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant unit ▶ Clean the assistant element with a soft cloth and a mild, non-abrasive clean‐ ing solution. 2.5.2 Sterilisation assistant element CAUTION Damage caused by sterilisation.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit Note Sterilise the suction cannulas according to the manufacturer's instructions. 2.5.4 Reprocessing of sieve inserts and suction hoses Note Check the sieve inserts daily and after each surgery.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit 2.6.2 Preparation of the tumbler holder and filler CAUTION Water in the device. Material damage from leaking water upon actuation of the "Tumbler filler" or "Bowl flush"...
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit CAUTION Damage caused by thermodisinfection. ▶ Do not subject the tumbler holder to thermodisinfection. The tumbler filler can be subjected to thermodisinfection.
▶ With regard to the "Water block, compact" assembly kit, please note that no disinfection facility is installed in the unit, and take appropriate safeguards. KaVo recommends to use the "Water block DVGW with integrated water dis‐ infection facility in combination with KaVo OXYGENAL 6 (Mat.
During normal operation of the treatment centre, the continuous germ reduction function secures continuous germ reduction in the water conducting systems due to the automatic dosing of the disinfectant, KaVo OXYGENAL 6. The rinsing programme and the intensive germ reduction facilitate the automatic rinsing and germ reduction of the water-conducting system.
Page 25
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene function of the water-conducting system Preparation for rinsing programme and intensive germ reduction ▶ Position the disinfection kit on the porcelain spittoon bowl.
Page 26
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene function of the water-conducting system ▶ Insert the triple-function or multifunction handpiece without cannulas into the clamping elements of the disinfection attachment. Make sure that the "W" but‐...
Page 27
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene function of the water-conducting system ▶ During the preparation for the rinsing program, press the foot control to set the time for the dental instrument rinsing and the triple-function or multifunction syringe between 0 seconds and 8:30 minutes.
During intensive germ reduction, the afferent water ducts are automatically filled with a higher concentration of KaVo OXYGENAL 6. The intensive germ reduction program lasts ca. 45 minutes. The increased hydrogen peroxide concentration acts in the system for at least 30 minutes.
Page 29
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene function of the water-conducting system Step Description Germ reduction of the triple-function and multifunctional handpiece and the spittoon bowl Duration: 60 seconds...
Page 30
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene function of the water-conducting system ▶ Press and hold down the "Motor anticlockwise direction" button, the "Tumbler filler" button and the foot pedal simultaneously.
The unit is now ready for use. Additional intensive germ reduction during period of non-use KaVo recommends carrying out an additional intensive germ reduction at univer‐ sities during semester breaks. This flushes the parts of the unit and replaces the germ reduction liquid.
Care instructions KaVo Primus® 1058 S/TM/C 2 Setup methods according to DIN EN ISO 17664 | 2.10 Preparation of the amalgam separator 2.9.2 HYDROclean function The HYDROclean function is an automatic programme for hydro-mechanical cleaning of the drainage and separation systems within and outside the treatment unit.
Note Replenish KaVo OXYGENAL 6 only if the acoustic signal is issued every 10 sec‐ onds. The signal can be turned off as follows: Hold down the control pedal, and move the backrest up and down. The next time the treatment unit is turned on, the signal sound is issued again.
Care instructions KaVo Primus® 1058 S/TM/C 3 Replenish and check the disinfectant | 3.1 Replenish OXYGENAL 6 ▶ Fill-up with KaVo OXYGENAL 6 (max. 1 litre). Note Do not fill through the centre opening (tumbler drain). ▶ Quit replenishing immediately if you hear a signal sound being issued every second (10 sounds).
Page 35
Note The water in the water bottle needs to be changed at least one daily. In order to ensure the efficiency of the measures, KaVo urgently recommends reprocessing the water with KaVo OXYGENAL 6. ▶ For cleaning, rinse the water bottle with water at less than 50 °C.
Page 36
In response to over-filling, the excess quantity above the tick mark is aspirated automatically to ensure exact dosing. ▶ Place the water bottle in the KaVo OXYGENAL 6 bottle. Make sure that the spout of the KaVo OXYGENAL 6 dosing facility (nose) is situated right above the opening of the water bottle.
Care instructions KaVo Primus® 1058 S/TM/C 3 Replenish and check the disinfectant | 3.2 Check the OXYGENAL 6 concentration ▶ Then fill-up the water bottle with tap water. Note Substances remaining in the dosing head are not contamination, but rather agents that are deposited there during extended periods of non-use after con‐...