KaVo MASTERtorque LUX M9000 LS Instructions For Use Manual
KaVo MASTERtorque LUX M9000 LS Instructions For Use Manual

KaVo MASTERtorque LUX M9000 LS Instructions For Use Manual

Hide thumbs Also See for MASTERtorque LUX M9000 LS:

Advertisement

Quick Links

Instructions for use
MASTERtorque LUX M9000 L – 1.008.7900
MASTERtorque LUX M9000 LS – 1.008.5400

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MASTERtorque LUX M9000 LS and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo MASTERtorque LUX M9000 LS

  • Page 1 Instructions for use MASTERtorque LUX M9000 L – 1.008.7900 MASTERtorque LUX M9000 LS – 1.008.5400...
  • Page 2 Manufacturer: Distributed by: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 Table of contents Table of contents 1 User instructions ....................... 2 Safety..........................2.1 Infection hazard ......................8 2.2 Improper use ........................ 8 2.3 Technical condition ......................9 2.4 Accessories and combination with other equipment............
  • Page 4 Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 Table of contents 9 Terms and conditions of warranty ..................30 4 / 32...
  • Page 5: User Instructions

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
  • Page 6 Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 1 User instructions Serial number Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity)
  • Page 7 Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 1 User instructions CAUTION In cases which – if not prevented – can lead to material damage. 7 / 32...
  • Page 8: Safety

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
  • Page 9: Technical Condition

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 2 Safety | 2.3 Technical condition 2.3 Technical condition If damaged, the device or components could injure patients, users and third parties. 4 Only operate devices or components if they show no signs of damage on the outside.
  • Page 10 Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 2 Safety | 2.6 Service and repair 4 After servicing, interventions on and repairs of the device and before re-use, have the service personnel perform safety checks on the device.
  • Page 11: Description Of The Product

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 3 Description of the product | 3.1 Intended use 3 Description of the product MASTERtorque LUX M9000 L (Mat. no. 1.008.7900) MASTERtorque LUX M9000 LS (Mat. no. 1.008.5400) 3.1 Intended use...
  • Page 12: Transportation And Storage Conditions

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 3 Description of the product | 3.3 Transportation and storage conditions M9000 L M9000 LS Recommended drive 2.8 bar (41 psi) 2.7 bar (39 psi) pressure Return air pressure <...
  • Page 13: Startup And Shut Down

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 4 Startup and shut down | 4.1 Installing the MULTIflex coupling in the M9000 L 4 Startup and shut down WARNING Hazard from contaminated products.
  • Page 14: Checking The Amount Of Water

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 4 Startup and shut down | 4.2 Checking the amount of water 4 Turn the spray ring clockwise to reduce the water volume. 4 Turn the spray ring anticlockwise to increase the water volume.
  • Page 15: Checking The Pressures, M9000 Ls

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 4 Startup and shut down | 4.4 Checking the pressures, M9000 LS A higher drive pressure will be reduced automatically by the medical device. 4.4 Checking the pressures, M9000 LS CAUTION Damage from soiled and moist cooling air/compressed air.
  • Page 16: Operation

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.1 Attaching the medical device, M9000 L 5 Operation NOTE At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 minutes (without transmission handpieces being at- tached) and if there is a risk of contamination from reflux or back suction, the system may also need to be rinsed for 20 to 30 seconds after each patient.
  • Page 17: Attaching The Medical Device, M9000 Ls

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.2 Attaching the medical device, M9000 LS 5.2 Attaching the medical device, M9000 LS WARNING Detaching the medical device during treatment. A medical device that is not properly locked can detach from the coupling dur- ing treatment.
  • Page 18: Removing The Dental Bur

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.5 Removing the dental bur CAUTION Dental bur with damaged, worn or deformed shafts. Injury hazard, dental bur may fall out during treatment.
  • Page 19 Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.5 Removing the dental bur CAUTION Damage to the chucking system. Material damage. 4 Never push the press-button while the dental bur is rotating.
  • Page 20: Troubleshooting

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 6 Troubleshooting | 6.1 Changing the O-rings on the MULTIflex coupling in the M9000 L 6 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts.
  • Page 21: Changing The Water Filter

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 6 Troubleshooting | 6.4 Changing the water filter CAUTION Overheating of the tooth due to insufficient amount of cooling water. Insufficient spray water can cause the medical device to overheat and damage the pulp and tooth.
  • Page 22: Reprocessing Steps In Accordance With Iso 17664

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use WARNING Hazard from contaminated products.
  • Page 23: Manual External Disinfection

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing 4 Remove the medical device from the spray attachment and let the cleanser act for 1 minute.
  • Page 24: Manual Drying

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing CAUTION Never reprocess the medical device with chloride-containing products. Malfunction and material damage.
  • Page 25: Automated Internal And External Cleaning And Internal And External Disinfection

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing CAUTION Never reprocess this medical device in an ultrasonic cleaner.
  • Page 26: Servicing With Kavo Spray

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing 7.4.1 Servicing with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use, i.e.
  • Page 27: Packaging

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging NOTE Close the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front flap and start the service procedure.
  • Page 28: Storage

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.7 Storage 7.7 Storage Reprocessed products must be stored appropriately protected from dust in a dry, dark and cool space with a low germ level.
  • Page 29: Optional Aids And Consumables

    Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. Spare turbine handpiece without wrench 2.000.2266 Wrench for cover 0.411.3053...
  • Page 30 Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...