Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 Table of contents Table of contents 1 User instructions ....................... 2 Safety..........................2.1 Infection hazard ......................8 2.2 Improper use ........................ 8 2.3 Technical condition ......................9 2.4 Accessories and combination with other equipment............
Page 4
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 Table of contents 9 Terms and conditions of warranty ..................30 4 / 32...
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
Page 6
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 1 User instructions Serial number Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity)
Page 7
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 1 User instructions CAUTION In cases which – if not prevented – can lead to material damage. 7 / 32...
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 2 Safety | 2.3 Technical condition 2.3 Technical condition If damaged, the device or components could injure patients, users and third parties. 4 Only operate devices or components if they show no signs of damage on the outside.
Page 10
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 2 Safety | 2.6 Service and repair 4 After servicing, interventions on and repairs of the device and before re-use, have the service personnel perform safety checks on the device.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 3 Description of the product | 3.1 Intended use 3 Description of the product MASTERtorque LUX M9000 L (Mat. no. 1.008.7900) MASTERtorque LUX M9000 LS (Mat. no. 1.008.5400) 3.1 Intended use...
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 3 Description of the product | 3.3 Transportation and storage conditions M9000 L M9000 LS Recommended drive 2.8 bar (41 psi) 2.7 bar (39 psi) pressure Return air pressure <...
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 4 Startup and shut down | 4.1 Installing the MULTIflex coupling in the M9000 L 4 Startup and shut down WARNING Hazard from contaminated products.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 4 Startup and shut down | 4.2 Checking the amount of water 4 Turn the spray ring clockwise to reduce the water volume. 4 Turn the spray ring anticlockwise to increase the water volume.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 4 Startup and shut down | 4.4 Checking the pressures, M9000 LS A higher drive pressure will be reduced automatically by the medical device. 4.4 Checking the pressures, M9000 LS CAUTION Damage from soiled and moist cooling air/compressed air.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.1 Attaching the medical device, M9000 L 5 Operation NOTE At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 minutes (without transmission handpieces being at- tached) and if there is a risk of contamination from reflux or back suction, the system may also need to be rinsed for 20 to 30 seconds after each patient.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.2 Attaching the medical device, M9000 LS 5.2 Attaching the medical device, M9000 LS WARNING Detaching the medical device during treatment. A medical device that is not properly locked can detach from the coupling dur- ing treatment.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.5 Removing the dental bur CAUTION Dental bur with damaged, worn or deformed shafts. Injury hazard, dental bur may fall out during treatment.
Page 19
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 5 Operation | 5.5 Removing the dental bur CAUTION Damage to the chucking system. Material damage. 4 Never push the press-button while the dental bur is rotating.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 6 Troubleshooting | 6.1 Changing the O-rings on the MULTIflex coupling in the M9000 L 6 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 6 Troubleshooting | 6.4 Changing the water filter CAUTION Overheating of the tooth due to insufficient amount of cooling water. Insufficient spray water can cause the medical device to overheat and damage the pulp and tooth.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use WARNING Hazard from contaminated products.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing 4 Remove the medical device from the spray attachment and let the cleanser act for 1 minute.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing CAUTION Never reprocess the medical device with chloride-containing products. Malfunction and material damage.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing CAUTION Never reprocess this medical device in an ultrasonic cleaner.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing 7.4.1 Servicing with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use, i.e.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging NOTE Close the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front flap and start the service procedure.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 7 Reprocessing steps in accordance with ISO 17664 | 7.7 Storage 7.7 Storage Reprocessed products must be stored appropriately protected from dust in a dry, dark and cool space with a low germ level.
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. Spare turbine handpiece without wrench 2.000.2266 Wrench for cover 0.411.3053...
Page 30
Instructions for use MASTERtorque LUX M9000 L – 1.008.7900, MASTERtorque LUX M9000 LS – 1.008.5400 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...