Instructions for use Globus 1078 S Table of contents Table of contents Table of contents User instructions User guide 1.1.1 Abbreviations 1.1.2 Symbols 1.1.3 Target group Service Warranty terms and conditions Transportation and storage 1.4.1 Currently valid packaging ordinance 1.4.2 Damage in transit 1.4.3...
Instructions for use Globus 1078 S Table of contents Setting functions 4.7.1 Select memory level Dentist 1 or Dentist 2 4.7.2 Fill the tumbler and rinse the spittoon 4.7.3 Turn the x-ray image viewer on and off 4.7.4 Activate bell 4.7.5...
Additional information can be obtained at: www.kavo.comAdditional information can be obtained at: www.kavo.com 1.3 Warranty terms and conditions KaVo provides the end customer with a warranty that the product cited in the hand‐ over certificate will function properly and guarantees zero defects in the material or 3/75...
Claims from this warranty can only be asserted when the transfer form (copy) be‐ longing to the product has been sent to KaVo, and the original can be presented by the operator or user.
ADSp. Art. 28). Outside of Germany Note KaVo is not liable for damage arising from transportation. Immediately inspect the delivery after receipt! If the packaging is visibly damaged on delivery, please proceed as follows: 1. The recipient of the package must record the loss or damage on the delivery receipt.
Page 8
Instructions for use Globus 1078 S 1 User instructions | 1.4 Transportation and storage Transport upright with the arrows pointing upwards Fragile - protect against impact! Protect from moisture. Permissible stacking load Temperature range Humidity Air pressure 6/75...
The user must ensure that that the unit works properly and is in a satisfactory con‐ dition before each use. The KaVo Globus 1078 S is a dental treatment unit in accordance with ISO 7494 that has a dental patient chair in accordance with ISO 6875. This KaVo product is...
Page 10
The overarching guidelines and/or national laws, national regulations and the rules of technology applicable to medical devices for start-up and use of the KaVo product for the intended purpose are to be applied and complied with.
Page 11
This must be done in strict com‐ pliance with all applicable national regulations. Questions on proper disposal of the KaVo product can be answered by the KaVo branch. Disposal of electronic and electrical devices Note According to EC directive 2002/96 concerning used electrical and electronic devi‐...
2.2.2 Product-specific Use and target group The KaVo Globus 1078 S is for treating children and adults in a dental context. The device may only be used by medical professionals. For technician jobs that require greater pressure than when working in the oral cavity such as grinding prostheses, etc., a special technician's machine must be used.
Instructions for use Globus 1078 S 2 Safety | 2.3 Safety instructions CAUTION Insufficient hygiene. Transmission of pathogens. ▶ To maintain the hygienic system, observe the care and servicing instructions. WARNING Injury or damage from damaged functional parts. When functional parts are damaged, it can cause additional damage or personal injury.
Page 14
Instructions for use Globus 1078 S 2 Safety | 2.3 Safety instructions CAUTION Germ formation. Infections. ▶ Before starting, rinse all the water drain lines without instruments. ▶ Before startup and after the device has not been used for a while (weekends, holidays, vacations, etc.), rinse or purge the air and water lines.
Page 15
Instructions for use Globus 1078 S 2 Safety | 2.3 Safety instructions CAUTION Danger of injury from overload. The patient chair can collapse. ▶ Do not subject the patient chair to a load above its limit (135 kg). CAUTION Hair may become caught in the patient's chair when the headrest is moved.
Instructions for use Globus 1078 S 3 Product description | 3.1 Treatment unit 3 Product description 3.1 Treatment unit ① Standard headrest ⑤ Armrest (extra) ② SOMATIC headrest ⑥ Chair base ③ 2-joint headrest ⑦ Comfort backrest ④ Seat ⑧ Progress backrest...
Instructions for use Globus 1078 S 3 Product description | 3.2 Unit base 3.2 Unit base ① Pivotable swing arm for holding the ④ Main fuse patient unit, device body and assis‐ tant unit ② Supply unit, customer connection of ⑤...
Instructions for use Globus 1078 S 3 Product description | 3.3 Device body with patient unit 3.3 Device body with patient unit ① Patient unit ④ Device body The central control is in the device body. ② Tumbler filler ⑤ Pressurised water bottle (extra) ③...
Instructions for use Globus 1078 S 3 Product description | 3.4 Dentist unit 3.4 Dentist unit ① Small x-ray viewer ⑥ Alternately use the INTRA LUX Motor K 200 or INTRA LUX Motor KL 701 ② Memospeed, Memodent ⑦ Keys and display field ③...
Instructions for use Globus 1078 S 3 Product description | 3.5 Assistant unit 3.5 Assistant unit ① Three-function handpiece or multi‐ ④ Saliva ejector functional handpiece ② Spray mist suction ⑤ Satelec Mini LED ③ Key and display field 18/75...
Instructions for use Globus 1078 S 3 Product description | 3.6 Key fields 3.6 Key fields The key fields on the dentist unit and assistant unit are the same. ① Patient chair ② function keys Group of keys for the patient chair...
Instructions for use Globus 1078 S 3 Product description | 3.7 Footswitch Designation Display LED Tumbler filler Cold light (on instruments)/ green Treatment unit ON when the instruments are moun‐ Spray preselection (on re‐ green/yellow moved instruments) Counterclockwise motor rotation 3.7 Footswitch...
Instructions for use Globus 1078 S 3 Product description | 3.8 Technical Specifications 3.8 Technical Specifications Drilling template and setup plan Drilling template (paper) Page 001 (Mat. no. 1.000.3001) Drilling template (sheet met‐ al)Mat. no. ). 10.002492 Setup plan (Mat. no. 1.000.3001)
Page 24
>45 mbar, recommended: 60 mbar Suction vacuum flow 500 Nl/min. The values apply to the KaVo measuring set (Mat. no. 0.411.8500) Operating environment Floor quality The quality of the floor must comply with the load bearing capacity for buildings DIN 1055 page 3 and the pressure re‐...
Instructions for use Globus 1078 S 3 Product description | 3.9 Rating plate and identification plate Transportation and storage conditions Ambient temperature -20 to +55°C Relative humidity 5% to 95% non-condensing Air pressure 700 to 1,060 hPa Weight Treatment unit 282 kg gross ,242 kg net 3.9 Rating plate and identification plate...
Page 26
Instructions for use Globus 1078 S 3 Product description | 3.9 Rating plate and identification plate Rating plate serial number Read and note the content of accompanying documents. Classification The handpieces are type BF application parts The patient chair is a type B application part...
Page 27
Instructions for use Globus 1078 S 3 Product description | 3.9 Rating plate and identification plate CE mark according to EC Directive 93/42 for medical devices VDE mark DVGW ID (Deutscher Verein des Gas- und Wasserfaches e.V.) (depending on the equipment)
Instructions for use Globus 1078 S 4 Operation | 4.1 Turn the unit on and off 4 Operation 4.1 Turn the unit on and off The main switch has the following tasks: ▪ Plug in or disconnect all the pins of the unit's plug into the customer's power supply.
Instructions for use Globus 1078 S 4 Operation | 4.2 Change from a right-handed to left-handed design 4.2 Change from a right-handed to left-handed design CAUTION Collision and damage to the unit. ▶ Ensure sufficient space for swinging. ▶ Turn off the device before conversion.
Page 30
Instructions for use Globus 1078 S 4 Operation | 4.2 Change from a right-handed to left-handed design ▶ Remove the seat. ▶ Carefully place the foot control in the spittoon bowl. ▶ Use protective base. ▶ Remove the cover ①.
Page 31
Instructions for use Globus 1078 S 4 Operation | 4.2 Change from a right-handed to left-handed design ▶ Lift the lock ② and simultaneously rotate the support arm toward the foot end. CAUTION Let the instrument hose hang on the patient chair or another object.
Page 32
Instructions for use Globus 1078 S 4 Operation | 4.2 Change from a right-handed to left-handed design ▶ Remove the lock ③, and simultaneously rotate the device body toward the backrest. 30/75...
Page 33
Instructions for use Globus 1078 S 4 Operation | 4.2 Change from a right-handed to left-handed design ▶ The support arm and device body are fixed when they audibly lock in place. ▶ Swing the dentist's unit into position, and rotate the patient's unit ① 360°.
Page 34
Instructions for use Globus 1078 S 4 Operation | 4.2 Change from a right-handed to left-handed design WARNING Safety lever not changed over. Collision of the armrests with the device body. ▶ Change over the safety lever. ▶ Change over the safety lever ①.
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair 4.3 Adjusting the patient chair 4.3.1 Swing the armrest (extra) To make it easier to get in and out, the armrest can be swung outward. ▶ Lift the armrest upward and swing it outward.
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair 4.3.3 Adjust head rest Standard headrest CAUTION Adjusting the headrest. Injury of neck muscles. ▶ Make sure that the patient is aware of the headrest setting.
Page 37
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair Swing the headrest ▶ Push the catch up ②, and swing the headrest into position. ▶ To lock the headrest, apply slight pressure (arrowA). SOMATIC headrest CAUTION Adjusting the headrest.
Page 38
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair Adjust the brake for the lengthwise headrest shift ▶ Remove the cover ③. ▶ Adjust the braking force using hexagon socket screw WAF 3 mm. ▶ Remount the cover ③.
Page 39
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair 2-joint headrest CAUTION Adjusting the headrest. Injury of neck muscles. ▶ Make sure that the patient is aware of the headrest setting. ▶ Patients need to raise their head slightly during adjustment.
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair Swing the headrest ▶ Push the lock ⑥ together, and swing headrest into the desired position. When swinging the headrest back into position, make sure that there is nothing in area B.
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair 4.3.5 Adjust chair position Gradually adjust the chair position Note The functions of the control keys on the dentist unit and assistant unit are the same.
Page 42
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair Note The seat and backrest can be moved at the same time. When the operating voltage for the patient chair is below 200 V, the movements in the automatic program can only be sequential.
Page 43
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair The chair positions can be saved on two footswitches; the standard setting is as follows: ▪ "Spray default" footswitch: "LP" automatic position (last position) ▪ "Blown air" footswitch: "SP" automatic position (rinsing position) The footswitch assignment can be changed.
Page 44
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair Retrieve the chair positions using the foot control Note If an instrument is removed, the chair functions of the foot control are blocked. The blocking can be removed by briefly pressing the stirrup switch. The functions are then available.
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair The chair moves into the automatic position. You can also use the cross switch. Lock the chair functions when the instrument is removed ▶ Press the foot pedal.
Page 46
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair The safety cut-offs can be found at the following places on the treatment unit. Check the patient chair and foot control safety shutoffs ① Backrest ④ Armrest ②...
Page 47
Instructions for use Globus 1078 S 4 Operation | 4.3 Adjusting the patient chair Safety shut-off on the patient's unit ① Patient unit swung over the patient chair The safety shutoff occurs went a movement angle has been exceeded, or part of the treatment unit collides with an object.
Instructions for use Globus 1078 S 4 Operation | 4.4 Move the dentist's unit 4.3.7 Special chair functions Note With operating voltages greater than 200 V, the spindle motors can operate simul‐ taneously, i.e., the chair and backrest can move simultaneously in the automatic program.
Instructions for use Globus 1078 S 4 Operation | 4.5 Adjust the height of the assistant unit 4.5 Adjust the height of the assistant unit The assistant's unit can be positioned vertically in three levels. ▶ To set a higher level, pull the assistant unit upward gently until it audibly locks in place.
Instructions for use Globus 1078 S 4 Operation | 4.7 Setting functions Note When the patient unit is swung over the patient chair, the safety shutoff is activated. 4.7 Setting functions 4.7.1 Select memory level Dentist 1 or Dentist 2...
Instructions for use Globus 1078 S 4 Operation | 4.7 Setting functions The spittoon is rinsed. ▶ To stop rinsing before the set time, press the "Spittoon" button again. 4.7.3 Turn the x-ray image viewer on and off ▶ Press the "X-ray viewer" button.
Instructions for use Globus 1078 S 4 Operation | 4.7 Setting functions Using the timer Setting the timer The minimum timer time is 30 seconds, and the maximum timer time is 8 minutes. ▶ Press the "Timer" button until you hear a beep.
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments 4.8 Using instruments The following sections describe the use and setting of the instruments. A distinction is drawn between operation with and without the installed Memospeed. The instruments are prevented from simultaneous use by software that determines when they are mounted.
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments ▶ Press the foot pedal several times when the instrument is mounted until the desired level is displayed. Memospeed Display ① Level display ② Time If stored values are changed during operation, these changed values will be lost when the unit is switched off and back on again.
Page 55
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments E123 ▶ Press the "Level selection" button until you hear a beep. ▶ To switch between maximum and minimum mode, press the "Increase value" or "Reduce value" button.
Page 56
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments ▶ "Preselected spray" footswitch. Two LEDs shine when the spray cooling level is activated. Storing values ▶ Press the "Level selection" button until you hear a beep. E123 The set values are saved for the set memory level and the set dentist level.
Page 57
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments Setting the rotational direction of the motor Note Only change the rotary direction of the motor when the motor is not running. ▶ Remove the motor from the holder.
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments The display switches to the setting menu for maximum. ▶ Press the "Increase value" or" Decrease value" button until the desired value is displayed. ▶ Press the "Preselect level" key.
Page 59
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments Note The torque wrench is subject to natural wear and should be changed when it longer works properly. (Mat. no. 1.000.4887) The following settings can be changed: ▪ Water emission ▪...
Page 60
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments E123 ▶ Press the "Preselect level" button until you hear a tone. ▶ Press the "Increase value" or" Decrease value" button until the desired value is set. Changing the high-pressure lamp CAUTION Injury hazard from hot valve body.
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments 4.8.6 Using the triple and multifunction handpiece ① "Air" button ③ "Water" button ② Cannula CAUTION Danger of injury from cannulas that are worn or not locked into place.
Page 62
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments Removing the cannulas ▶ Remove the cannula with a slight amount of rotation while holding the handpiece on the grip sleeve. Setting the cold-light intensity The cold light intensity is adjusted in the same manner as with the turbine.
Instructions for use Globus 1078 S 4 Operation | 4.8 Using instruments 4.8.7 Setting the spray mist suction device and saliva ejector CAUTION Unintentionally activating the saliva ejector or spray fog suction. Injury in the area of the mouth. ▶ When using the vacuum stop to interrupt suction, do not leave the saliva ejector or spray fog suction in the patient's mouth.
Instructions for use Globus 1078 S 5 Preparation methods DIN EN ISO 17664 5 Preparation methods DIN EN ISO 17664 Note The preparation methods can be found in the care instructions. 62/75...
Instructions for use Globus 1078 S 6 Adaptable accessories and kits | 6.1 Unit base 6 Adaptable accessories and kits 6.1 Unit base Designation Description DVGW water block Water block, compact Without DVGW Water bottle DVGW Intensive disinfection kit Steel support base...
Instructions for use Globus 1078 S 6 Adaptable accessories and kits | 6.4 Dentist unit 6.4 Dentist unit Designation Description Multiflex LUX hose For connecting the turbine and SONICflex. LUX motor hose and IN‐ For connecting the INTRA LUX motor K 200 and INTRA TRA K and INTRA KL mo‐...
Instructions for use Globus 1078 S 7 Troubleshooting 7 Troubleshooting Note In case of malfunctions, consult the separate instructions for the use and care of the individual instruments (such as the turbine, motor, camera, Satelec Mini LED, etc.). Malfunction Cause Remedy Nothing works.
Page 68
Instructions for use Globus 1078 S 7 Troubleshooting Malfunction Cause Remedy Sieves in suction connector ▶ Exchange the sieve. blocked. Foot switch for vacu-stop has been ▶ Release the foot switch. pressed. Suction machine not running. ▶ Turn on the suction machine.
▪ Type BF comment 2 ▪ Measurement according to EUL/EPA KaVo offers a medical device book for keeping an inventory and recording essential master data on the medical device. The medical device book is only required in Germany and is therefore only in German (Mat. no. 0.789.0480).
Page 70
Instructions for use Globus 1078 S 8 Do safety checks | 8.1 Measure protective conductor resistance Note Take into account the additional measuring points of additional equipment such as connection to equipment by a different manufacturer, treatment lamps, multimedia system, etc.
Instructions for use Globus 1078 S 8 Do safety checks | 8.2 Measure equivalent unit leakage current 8.2 Measure equivalent unit leakage current Threshold: < 10 mA The equivalent unit leakage current (EUL) can be measured using an EPA meas‐...
Instructions for use Globus 1078 S 8 Do safety checks | 8.3 Measure equivalent patient leakage current Note Take into account the additional measuring points of additional equipment such as connection to equipment by a different manufacturer, treatment lamps, multimedia system, etc.
Page 73
Instructions for use Globus 1078 S 8 Do safety checks | 8.3 Measure equivalent patient leakage current Note Take into account the additional measuring points of additional equipment such as connection to equipment by a different manufacturer, treatment lamps, multimedia system, etc.
Page 74
Instructions for use Globus 1078 S 8 Do safety checks | 8.3 Measure equivalent patient leakage current ▶ Check the EPA at the measuring points or with an EPA measuring line Mat. no. 1.001.9904. 72/75...
9.1 Electromagnetic Transmissions The Globus 1078 S treatment unit is for use in an environment like the one cited below. The customer or user of the Globus 1078 S should ensure that it is used in the correct environment. Measurements of noise transmis‐...
The Globus 1078 S is intended for use in an electromagnetic environment in which HF disturbances are controlled. The customer or the user of the Globus 1078 S can help prevent electromagnetic disturbances by maintaining the minimum distance...
Page 77
Instructions for use Globus 1078 S 9 Data on electromagnetic compatibility according to EN60601-1-2 | 9.3 Recommended safe distance between portable and mobile HF telecommunications equipment and the treatment unit NOTE 2: These guidelines may not be applicable in every case. The spread of electromagnetic waves is absorbed and reflected by buildings, objects and people.