Enter Check Standard Values
Each standard has two values assigned, for two individual
wavelengths.
Use the up/down keys to adjust the values to match the
certificates, following the order defined on the display. Press
OK when the correct value is shown and the prompt will
forward to the next value.
Upon completion the message Check Standard Values
assigned successfully will be displayed.
Press OK to return to the Check Standard Mode menu.
Check Standard Measurement
Follow the on-screen prompts to insert the Check Standards
in the defined order. The Photometer 7500 Bluetooth will
automatically measure the Transmittance at the required
wavelength.
Upon completion of the sequence the results are displayed on
screen with pass or fail status.
If the Check Standard Mode reports a failure, see Appendix 2
Troubleshooting for guidance or contact your local Palintest
supplier.
36