Multi-Point Measurement; Items Required; Measuring Method - Konica Minolta T-10A Instruction Manual

Illuminance meter
Table of Contents

Advertisement

Multi-Point Measurement

This instrument allows you to perform multi-point measurement by using more than two optional
receptor heads and adapters. (Up to 30 receptor heads and adapters can be connected.)
Since each receptor head must be powered constantly during multi-
point measurement, the optional AC adapte must be used.

Items Required

z Main body adapter
(with an extension cable) : T-A20 (optional)
z Receptor head adapter
z AC adapter
When the number of receptor heads is 10 or less
When the number of receptor heads is 30 or less
If the cable supplied with the adapter is not long enough:
z Use a commercially available 10Base-T network cable (category 5 straight cable). The cable can be
extended up to 50 meters. Using a cross cable will disable the instrument and may cause malfunction
of the connected system.

Measuring Method

z If measurement is performed with the receptor head connected to the main body with the extension
cable, the measured values and operation of the instrument may be affected if equipment which
generates electrical noise is present near the extension cable. In this case, such equipment must be
kept away from the cable during measurement.
: T-A21 (optional)
: AC-A308 (optional)
: AC-A311 (optional)
1
Set the power switch to "O" (OFF), and attach the AC adapter
to the instrument.
2
Attach the main body adapter to the instrument.
z The receptor of T-10A or mini receptor of T-10MA is not
compatible with existing model T-10 or T-10M.
41