Page 1
ILLUMINANCE METER T-10A/T-10MA INSTRUCTION MANUAL Before using this instrument, please read this manual.
Page 2
Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incorrect use of the instrument. Denotes a sentence regarding a safety warning or note. Read the sentence carefully to ensure safe and correct use. Denotes a prohibited operation.
WARNING the instrument or other property.) Do not use batteries other than those specified by KONICA MINOLTA. When installing batteries in the instrument, make sure that they are correctly oriented according to the (+) and (-) marks. Also make sure not to mix new and old batteries, or mix batteries of different types. Failure to adhere to these instructions may cause batteries to explode or leakage of electrolyte, resulting in fire, injury or air pollution.
Page 4
-20 and 55 °C at instrument in these types of locations. 85% or less relative humidity (at 35 °C). Do not z The receptor of T-10A or mini receptor store this instrument in hot and humid areas of T-10MA is not compatible with or areas where condensation is likely to occur.
Contents SAFETY PRECAUTIONS ........................ 1 Contents ............................3 Names and Functions of Parts ...................... 4 Basic Operations Preparation ............................. 8 Installing the Receptor Head..........................8 Removing the Receptor Head ..........................8 Selecting measuring units ..........................8 Attaching the Cap ..............................8 Putting the Instrument in the Case ........................10 Installing the Batteries .............................11 Turning Power ON ............................11 Zero Adjustment ..............................12...
Names and Functions of Parts T-10A Mode selector keys ▲When the slide cover is open 1 Receptor Window 2 Display Mode selector keys 3 [NORM] key ......Switches to normal illuminance measurement mode. 4 [⊿] key ........Switches to illuminance difference/ratio measurement mode, and switches the displayed value between the illuminance difference and ratio.
Page 7
11 [SET] key ......• Activates SET mode. • Confirms the numeric setting. 12 [▲][▼] keys ......Changes the currently set value and decimal point position. 13 [ /D-OUT] key ....In normal mode, this key is used to output data to a personal computer or printer.
Page 8
T-10MA Mini receptor head Connecting plug Body no. Receptor window Notes on T-10MA Body No. A body No. is indicated in two places as shown below: one on the sensor (receptor head code) and the other on the back of the adapter (main body side). When using two or more T-10MA (mini receptor head) units, make sure that the body no.
ON may generate an error. Selecting measuring units The Konica Minolta llluminance Meter allows you to take readings in either Lux (lx.) or Foot-candle (fcd) units. Set the power switch to “ ” (OFF), and pull the battery cover while pressing it down slightly.
Page 11
If the strap is not to be used, install the cap using the procedure as shown below. Untie the strap attached to the cap, and remove the strap from the cap. Attach the strap to the strap hook as shown. Pass the strap through the cap, and knot the end of the strap so that it does not become detached from the cap.
Putting the Instrument in the Case ■ For T-10A Attach the cap to the receptor window and put the instrument into the case. ■ For T-10MA Remove the sensor plug from the mini receptor head, put the instrument into the case, then put the sensor into the pocket of the case.
Installing the Batteries WARNING Do not allow batteries to come into contact with fire, and do not charge them (unless they are specifically designated as rechargeable), heat them, or take them apart. Any of these actions may cause batteries to overheat or explode, possibly resulting in fire or injury. CAUTION Do not use battery types that are not specified for this instrument.
Zero Adjustment Zero adjustment is performed automatically when the power switch is set to “I” (ON). z “CAL” (calibration) will be displayed in the display section during zero adjustment. z Zero adjustment is performed electrically, so no cap is required. z When zero adjustment is complete, “CAL”...
Measuring the Illuminance z When performing measurement, take care not to allow the shadow or reflection of the operator to enter the receptor window. Measuring Method Press the [NORM] key. Push out the hold button (RUN). Place the receptor head in the desired measuring position. ...
Measuring the Illuminance Difference / Ratio This section explains how to measure the difference between the measured illuminance and reference illuminance as well as the ratio of the measured illuminance to the reference illuminance. Entering the Reference Value The reference value can be set in the following two ways. j By setting the measured value as the reference value k By entering the desired reference value numerically Setting the measured value as the reference value...
Page 17
Open the slide cover and press the [SET] key. The reference value which has been held will blink. z To cancel the reference value, press the [CALL] key. Press the [SET] key again. The reference value will be set, and “± 0 lx” or “100.0 %” will appear. while the z To check the reference value, press the [CALL] key.
Page 18
Open the slide cover and press the [SET] key. The instrument is now ready for you to enter a value, and one of the following values will appear. z If no reference value had been set, “00.00” will appear (with the decimal point blinking). z If the reference value had already been set, it will appear (with the decimal point, “0”...
Page 19
Press the [ ► /D-OUT] key to select the digit to be changed. The value at the selected digit will blink. Change the value using the [ ▲ ] and [ ▼ ] keys. Repeat steps 5 and 6 to complete entry of the desired reference value.
Measuring the Illuminance Difference / Ratio Push out the hold button to switch from HOLD to RUN. Place the receptor head in the desired measuring position. The measured illuminance difference or ratio will appear. ⊿ z Each time the [ ] key is pressed, the illuminance difference or ratio will appear alternately.
Measuring the Integrated Illuminance This instrument enables you to measure integrated illuminance (lx·h), integration time (h) and average illuminance (lx). Set the power switch to “O” (OFF) and push in the hold button (HOLD). Set the power switch to “I” (ON). z “0 lx”...
Page 22
Displays the integrated illuminance. → ↓ Displays the integration time. ↓ Displays the average illuminance. z Integration can be continued even if measurement of illuminance or illuminance difference/ be paused. z To reset integration, set the power switch to “O” (OFF). z The maximum measurable value and minimum unit for each display mode are given in the table below.
Automatic Integration Stop Function The instrument has a function that stops measurement automatically when the preset integration time or integrated illuminance is reached during measurement of integrated illuminance. z It is not possible to preset both integration time and integrated illuminance at the same time. The integration time or integrated illuminance, whichever is set last, will be effective.
Page 24
Open the slide cover and press the [SET] key. The instrument is now ready for you to enter a value, and “00.00” will appear (with the decimal point blinking). z If the value (integration time or integrated illuminance) had already been set, it will appear (with the decimal point, “0”...
Page 25
▼ Integration time (0.001〜9.999) → ↓ (0.01〜99.99) ↓ (0.1〜999.9) ↓ (1〜9999) * In the above example, the [ ▲ ] key is pressed to change the display. The display will change in opposite direction if the [ ▼ ] key is pressed. * Values in ( ) indicate the settable range.
The unit will blink, indicating that measurement is currently paused. z This also occurs in normal and ⊿ modes. Set the power switch to “O” (OFF). z If the power switch is not set to “O” (OFF) to reset the instrument, it will also be impossible to perform measurement in other modes.
Doing so may result in a fire. In such situations, turn the power OFF immediately, remove the batteries (or disconnect the AC adapter from the AC outlet), and contact the nearest Konica Minolta-authorized service facility. CAUTION The voltage/ampere rating of the AC adapter jack’s is 8 V DC / 1.5 A.
When measuring under a light source which has a considerably different spectral distribution from T-10A/T-10MA’s calibration light source, the instrument will cause an indication error due to a slight deviation of the relative spectral response from spectral luminous efficiency (Vλ). To correct this error, this instrument has CCF function (mode), allowing you to set the color correction factor (CCF).
CCF Setting Method The CCF can be set in the following two ways. Once the CCF is set, it will remain effective even if the power is turned OFF or the batteries are changed. j Entering the CCF directly This method can be used if the CCF is known. ■...
Page 32
Change the CCF using the [ ▲ ] and [ ▼ ] keys. Each time the [▲] or [ ▼ ] key is pressed, the value will change in 0.001 steps (settable range: 0.500 to 2.000). Holding the [▲] or [ ▼ ] key will cause the displayed value to change continuously.
Page 33
Press the [SET] key. The instrument is now ready for you to enter a value, and the currently displayed measured value will blink. Enter the value using the [ ▲ ] and [ ▼ ] keys. Each time the [▲] or [ ▼ ] key is pressed, the value will change in steps of the smallest digit.
Setting the Range Manually The measuring range is switched automatically from one to another during measurement (5 ranges are available). When you want to fix the range or you want to perform measurement of illuminance continuously using the analog output function (see page 28), set the measuring range as follows. Manual/auto measuring ranges Measuring range (unit: lx) Measuring range (unit: fcd)
Open the slide cover and select the desired range using the [ ▲ ] and [ ▼ ] keys. The range will be set immediately when it is selected. → z Each time the [▲] or [ ▼ ] key is pressed, the range switches from one to another in the following order and the ↓...
Recording the Measured Illuminance Continuously (Analog Output) Have the instrument ready for operation as described in “Measuring the illuminance” (page 13) and push out the hold button (RUN). z The range can be switched in RUN state only. Insert the Mini plug (optional) into the analog output terminal. z How to connect the Mini plug (optional) When connecting the instrument to an external device, solder the Mini plug as shown in the figure.
Page 37
Adjust the zero level of the recorder so that the analog output voltage from the instrument equals to the zero level of the recorder. Remove the cap (or block). An analog signal (i.e. voltage) proportional to the measured illuminance will be output, and the measured illuminance will be recorded continuously to the recorder.
Printing the Measured Data (Digital Output) Digital Output of the Measured Value This section explains how to print out the measured data. To print out the data, the printer cable T-A12 (optional) is required to connect a printer to the instrument. Printer Requirements The printer must meet the following requirements.
Set the instrument’s power switch to “I” (ON), then turn ON the printer. Push out the hold button (RUN), and start measurement. Press the [ ► /D-OUT] key. Each time the key is pressed, the data effective when the key is pressed will be sent to the printer and printed out.
If you install KONICA MINOLTA’s standard T-S10w data management software on your computer, you will also be able to use the computer to view measurement results in real time and to control multi-point measurements.
Set the power switch to “O” (OFF). Attach the main body adapter to the instrument. z The receptor of T-10A or mini receptor of T-10MA is not compatible with existing model T-10 or T-10M. ▼...
Connect both adapters with the extension cable. Set the power switch to “I” (ON) and push out the hold button (RUN). Measurement will start. z If the receptor head is connected to the main body through the extension cable and measurement is performed, consumption current will increase by approximately twice.
Set the power switch to “O” (OFF), and attach the AC adapter to the instrument. Attach the main body adapter to the instrument. z The receptor of T-10A or mini receptor of T-10MA is not compatible with existing model T-10 or T-10M.
Page 44
Attach a receptor head adapter to each receptor head (mini ▼For T-10MA receptor head in the case of T-10MA). Connect both adapters an extension cable. Connect the remaining receptor heads serially with extension cables. Set an ID no. to each receptor head using the rotary switch of the adapters.
Page 45
Set the power switch to “I” (ON). The smallest ID no. among those currently set will be displayed at the upper right of the display section. z Press the [▲] or [ ▼ ] keys to check that the ID no. switches from one to another.
Notes on Multi-Point Measurement Receptor Head ID No. z Be sure to set unique ID numbers for all heads. If you set the same number on more than one head, the ILLUMINANCE METER will fail recognizing the heads, resulting in measurement errors. Illuminance Difference/Ratio Measurement z The reference value must be set for each receptor head.
Standard Accessories Optional Accessories Batteries Main body adapter T-A20 Strap Receptor head adapter T-A21 z With an extension cable (1 m) Cap T-A13 Mini plug (for T-10 only) Mini Cap T-A14 Connecting cable (for T-10M only) T-A11 (for personal computer) T-A12 (for printer) Case T-A10 AC adapter...
Page 49
T-10A receptor head Data Management z With a cap Software T-S10w T-10MA receptor USB Cable head (mini type) T-A15 z With a mini cap T-10W A receptor head (water-proof mini type: 5 m) z With a mini cap T-10W A receptor...
Illuminance meter T-10W receptor head T-10W receptor head T-10MA T-10A Case T-10W *Custom order T-A10 T-10W *Custom order External power T-A13 AC Adapter for T-10A only AC-A308 AC-A311 Hard Case Mini Cap CL-A10 T-A14 for T-10MA only Standard Accessories Optional Accessories...
If this message reappears, because, for instance, the receptor head contact the nearest Konica Minolta- was not connected to the main body at authorized service facility. the time the power was turned ON.
(Unit: mm) The top of the receptor window is used as the reference plane for measurement as shown below. Center of receptor window Cable length T-10M:1m Reference plane Reference plane Center of receptor window φ25 Center of tripod socket T-10A T-10MA...
Ideally, the relative spectral responsivity of the illuminance meter should match V (λ) the spectral luminous efficiency of the human eye for photopic vision. As shown in the graph above, the relative spectral responsivity of Konica Minolta Illuminance Meters T-10A/T-10MA is within 6% (f ’) of the JIS spectral luminous efficiency V (λ).
Page 55
Ideally, the relative spectral responsivity of the illuminance meter should match V (λ) the spectral luminous efficiency of the human eye for photopic vision. As shown in the graph above, the relative spectral responsivity of Konica Minolta Illuminance Meters T-10A/T-10MA is within 6% (f’...
The graph above shows the cosine correction characteristics of Konica Minolta Illuminance Meters T-10. The cosine error of T-10 are shown in the table right.
Page 57
The graph above shows the cosine correction characteristics of Konica Minolta Illuminance Meters T-10. The cosine error of T-10 are shown in the table right.
Page 58
Specifications Illuminance Meter T-10A Illuminance Meter T-10MA Model (Standard receptor head) (Mini receptor head) Multi-function digital illuminance meter with detachable receptor head Type (Multi-point measurements of 2 to 30 points is possible) Conforms to requirements for Class AA of JIS C 1609-1: 2006 Illuminance meter class "Illuminance meters Part 1: General measuring instruments"...
Page 59
[CAUTION] KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE MISUSE, MISHANDLING, UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT, OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE USE OF OR INABILITY TO USE THIS PRODUCT.
Need help?
Do you have a question about the T-10A and is the answer not in the manual?
Questions and answers