Separating the Receptor Head from the Main Body

By using the optional adapters T-A20 (for main body) and T-A21 (for receptor head),
the receptor head can be separated from the main body for measurement.

Items Required

z Main body adapter
(with an extension cable)
z Receptor head adapter
If the cable supplied with the adapter is not long enough:
z Use a commercially available 10Base-T network cable (category 5 straight cable). The cable can be
extended up to 100 meters. Using a cross cable will disable the instrument and may cause malfunction
of the connected system.

Measuring Method

z If measurement is performed with the receptor head connected to the main body with the extension
cable, the measured values and operation of the instrument may be affected if equipment which
generates electrical noise is present near the extension cable. In this case, such equipment must be
kept away from the cable during measurement.
For T-10MA
: T-A20 (optional)
: T-A21 (optional)
1
Set the power switch to "O" (OFF).
2
Attach the main body adapter to the instrument.
z The receptor of T-10A or mini receptor of T-10MA is not
compatible with existing model T-10 or T-10M.
3
Attach the receptor head adapter to the receptor head (mini
receptor head in the case of T-10MA).
39