Reference Standard RS 2x30, RS 1x30
User's Guide
Commands for configuring SU/SI contacts behavior in AUTO mode:
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:STATe {<state>}
Setting of SU contact state in AUTO mode after start of dosage function. Parameter <state> is
0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:STATe?
Request for setting of SU contact state in AUTO mode after start of dosage function. Reply is
0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:STATe {<state>}
Setting of SI contact state in AUTO mode after start of dosage function. Parameter <state> is
0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:STATe?
Request for setting of SI contact state in AUTO mode after start of dosage function. Reply is 0(contact
OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:STATe {<state>}
Setting of SU contact state in AUTO mode before start and after end of dosage function. Parameter
<state> is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:STATe?
Request for setting of SU contact state in AUTO mode before start and after end of dosage function.
Reply is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:STATe {<state>}
Setting of SI contact state in AUTO mode before start and after end of dosage function. Parameter
<state> is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:STATe?
Request for setting of SI contact state in AUTO mode before start and after end of dosage function.
Reply is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:DELay {<delay>}
Setting of SU contact switch delay in AUTO mode after start of dosage function. Parameter <delay> is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:DELay?
Request for setting of SU contact switch delay in AUTO mode after start of dosage function. Reply is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:DELay {<delay>}
Setting of SI contact switch delay in AUTO mode after start of dosage function. Parameter <delay> is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:DELay?
Request for setting of SI contact switch delay in AUTO mode after start of dosage function. Reply is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:DELay {<delay>}
Setting of SU contact switch delay in AUTO mode before start and after end of dosage function.
Parameter <delay> is delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:DELay?
Request for setting of SU contact switch delay in AUTO mode before start and after end of dosage
function. Reply is delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:DELay {<delay>}
Setting of SI contact switch delay in AUTO mode before start and after end of dosage function.
Parameter <delay> is delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:DELay?
Request for setting of SI contact switch delay in AUTO mode before start and after end of dosage
function. Reply is delay in seconds (e.g. 0.05).
www.appliedp.com
49
RS-UG91-EN