Features And Functions; System Time; Display - Applied Precision RS 2130 User Manual

Reference standard, standard meter of electrical power and energy
Table of Contents

Advertisement

Device has its own clock and calendar - date and time can be used (set and red).
They can be controlled by menu item System / Date & Time or by remote interface commands:
SYSTem:TIMe?
Request for the device system time. Reply is the text string in "YYYY-MM-DD hh:mm:ss.xx" format.
SYSTem:TIMe {<time>}
Setting of the device system time to value defined by <time> parameter, which is a text string in
"YYYY-MM-DD hh:mm:ss.xx" format. Change in the device system time happens immediately.
SYSTem:TIMe:BEGin?
Request for the last switch-on (or reset) time of the device. Reply is the text string in "YYYY-MM-DD
hh:mm:ss.xx" format.
Definition of format "YYYY-MM-DD hh:mm:ss.xx" (example: "1996-05-20 17:23:19.04730"):
YYYY ....... year (4 digits)
MM .......... month (2 digits)
DD ........... day (2 digits)
Display functions and parameters can be controlled using these remote commands:

DISPlay?

Request for the setting of measurement results depicting on the device display.
If it is switched-off the reply is "0", otherwise is "1".
DISPlay {OFF|ON}
Switch off | switch on of the measurement results depicting on the device display. Switching-off means
empty display during the measurement. However, activated device menu is always displayed.
DISPlay:TEXT?
Request for the setting of the user text, which is depicted on the device display. The empty string means
user text extraction switch off.
DISPlay:TEXT {<"text">}
User text extraction (bordered by quotation marks) on the device display. Sending this command with
the empty string can do user text extraction switch off.
DISPlay:RESolution?
Request for state of Display Resolution mode. Answer is 0 (Standard), 1 (High) or 2 (Maximal). In High
and Maximal modes are measured results shown more precisely (with more decimal places).
DISPlay:RESolution {<state = 0 | 1 | 2>}
Setting of state of Display Resolution mode. Parameter is 0 (Standard), 1 (High) or 2 (Maximal). In High
and Maximal modes are measured results shown more precisely (with more decimal places).
DISPlay:LARGe?
Request for state of Large Font mode. Answer is 0 (disabled) or 1 (enabled).
If Large Font is enabled and only one unit is selected in Quantities submenu then measured results of
this unit are shown with large font.
DISPlay:LARGe {<state = 0 | 1>}
Setting of state of Large Font mode. Parameter is 0 (disabled) or 1 (enabled).
If Large Font is enabled and only one unit is selected in Quantities submenu then measured results of
this unit are shown with large font.
DISPlay:REFResh
Immediate refresh of display.
www.appliedp.com

5. Features and Functions

5.1 System Time

hh ...... hour (2 digits)
mm .... minutes (2 digits)
ss ...... seconds (2 digits)
xx ...... fraction of second (max. 5 decimal digits)
5.2 Display
Reference Standard RS 2x30, RS 1x30
RS-UG91-EN
User's Guide
21