Applied Precision RS 2130 User Manual page 31

Reference standard, standard meter of electrical power and energy
Table of Contents

Advertisement

SYSTem:ENERgy:IMPulse:ECONstant?
Request for the setting of the constant of the meter under test connected to Reference Standard via
external Local Evaluation Unit OPS. Answer is the constant value in unit 1/kWh or 1/kVAh.
SYSTem:ENERgy:IMPulse:ECONstant {<constant>}
Setting of the constant of the meter under test connected to Reference Standard via external Local
Evaluation Unit OPS which is using 10000 nominal impulses for evaluation of error. Parameter
<constant> is in unit of 1/kWh or 1/kVAh.
Setting this parameter changes also device impulse output constant parameter using this formula:
IMPulse = ECONstant * 10000 / DIVider.
SYSTem:ENERgy:IMPulse:DIVider?
Request for the impulse division factor, which can be used for dividing device impulse output constant
(and also constant of the meter under test) parameters.
SYSTem:ENERgy:IMPulse:DIVider {<divider>}
Setting of the impulse division factor, which can be used for dividing device impulse output constant
(and also constant of the meter under test) parameters. Default value of impulse division factor is 1.
SYSTem:ENERgy:DOSage:...
Remote interface commands for DOSAGE function. See chapter DOSAGE.
SYSTem:ENERgy:OUTx:STATe? x=0,1,2 or 3
Request for the state of the device impulse output OUTx. Reply is: 0 (for OFF), 1 (for ACTIVE POWER),
2 (for REACTIVE POWER), 3 (for FIXED) or 4 (for APPARENT POWER).
SYSTem:ENERgy:OUTx:STATe
Setting of the state of device impulse output OUTx. Parameter is state code (0 to 4 listed in previous
command) or state name.
SYSTem:ENERgy:OUTx:CHANnel?
Request for the setting of the individual channel selection for impulse output OUTx functions related to
the energy. Reply are three (in a sequence for channels L1, L2 and L3) values 0 or 1 (0 denounce for
switched-off and 1 for switched-on given channel selection) separated by a commas (e.g. "1,0,1").
SYSTem:ENERgy:OUTx:CHANnel {0|1|OFF|ON},{0|1|OFF|ON},{0|1|OFF|ON} x=0,1,2 or 3
Setting of the individual channel selection for impulse output OUTx functions related to the energy. The
parameters value (in a sequence for channels L1, L2 and L3) is ON or 1 for switched-on state and OFF
or 0 for switched-off state of given channel selection. This setting influences the summing up of the
individual channels into the value of the generated frequency on the device impulse output OUTx if this
is in the (ACTIVE, REACTIVE or APPARENT) ENERGY mode.
SYSTem:ENERgy:OUTx:FIXed?
Request for the constant frequency value (which is generated if the impulse output OUTx is in the state
FIXED). Reply is the text string containing a real number with the value of this frequency (e.g.
"+1.000000E+03"). This value is in a Hz unit.
SYSTem:ENERgy:OUTx:FIXed {<freq>}
Setting of the constant frequency value (which is generated if the impulse output OUTx is in the state
FIXED). Parameter <freq> is the frequency value in Hz unit.
SYSTem:ENERgy:OUTx:FACTor? x=0,1,2 or 3
Request for the setting of the multiplication constant on the device impulse output OUTx. Reply is real
number with the value of the set multiplication constant (e.g. "+2.500000E+02"). The unit of this
constant is 1/Ws or 1/VAs.
SYSTem:ENERgy:OUTx:FACTor {<factor>}
Setting of the multiplication constant of the device impulse output OUTx which generates the impulse
signal proportional to the measured energy. Parameter <factor> is the multiplication constant of the
impulse output in the unit of 1/Ws or 1/VAs.
www.appliedp.com
Reference Standard RS 2x30, RS 1x30
{0|1|2|3|4} | { OFF | ACTIVE | REACTIVE | FIXED | APPARENT }
x=0,1,2 or 3
x=0,1,2 or 3
x=0,1,2 or 3
x=0,1,2 or 3
RS-UG91-EN
User's Guide
x=0,1,2 or 3
31