Applied Precision RS 2130 User Manual page 36

Reference standard, standard meter of electrical power and energy
Table of Contents

Advertisement

th
6
Line:
- k ............................... meter constant of the meter under test (set in Measurement / Energy /
th
7
Line:
- AVG ......................... average value from all measures samples shown after getting second
th
8
Line:
- STD ......................... standard deviation value of all measures samples shown after getting
th
th
6
- 8
Line:
- samples .................. list of measured samples with their sequence numbers. List can be
Right part of window contains list of available keys – some of them are associated with parameters
which values are shown next to key indicator.
- 1 or F1 / Meas ......... key for entering or leaving MEAS mode of the OPS unit. This key starts /
- 2 or F2 / Calib ......... key for entering or leaving BARGRAPH / CALIB mode of the OPS unit.
- 3 or F3 / Input ......... key for setting input mode: LED, DISK or S0.
- 4 or F4 / Res. .......... key for setting resolution: LOW = results with two decimal places and
- 5 or F5 / Wind. ........ key for setting measurement window used by OPS during capturing marks
- 6 or F6 / Imp. .......... key for setting of "OPS predivider" e.g. number of impulses captured by
- 7 / Samp. ................ key for setting of number of samples needed for evaluation of test. Values
- 8 / Mode .................. key for setting testing mode: Meter Error Test, Meter Constant Test with
- 9 / Clr ...................... key for clearing of history (list) of measured samples. Usage of this key
36
Impulse submenu)
sample and updated after each new sample. After finishing of test
(measuring required number of samples) is this value result of test.
second sample and updated after each new sample.
Middle part of window
browsed using UP and DOWN arrows.
Right part of window
ends meter testing when all other parameters are set to correct values.
Actual state is indicated in end of line: ON or OFF.
Actual state is indicated in end of line: ON, SET or OFF.
10000 nominal impulses (100 ppm resolution) or HIGH = results with three
decimal places and 60000 nominal impulses (17 ppm resolution)
from rotation disks. Value can be 0-100%.
OPS unit needed for every sample. Value can be 1-60000.
can be 1-50.
value shown in [/kWh or /kVAh] or Meter Constant Test with value shown
in [/Ws or /VAs]
when testing is in progress makes reset of the test without interruption of
the test.
RS-UG91-EN
Reference Standard RS 2x30, RS 1x30
User's Guide