In a basic mode (when actual values of measured quantities are displayed) it is possible to change the
display contrast (without necessity to enter the device menu) using the following keys combinations:
1.
+/- (Func key) and UP
2.
+/- (Func key) and DOWN
3.
+/- (Func key) and LEFT
4.
+/- (Func key) and RIGHT
Remark: Display contrast change using keys is not possible on old version devices where this can be
done only by using a small potentiometer placed on the front panel.
Measurement....
Special func. ...
System ...
Calibration...
Save Config.
Def. Config.
Input / Range...
Energy ...
Frequency Mode ... - menu of output frequency settings,
Operation Mode
Time Base
Harm. Meas Mode
Harm. Meas Limit
Appar. Meas Mode - setting the way of apparent power acquisition. There are two options –
Vector Diagram ->
12
Immediate Display Contrast Change

3.2 Menu Description

Structure of the device menu is shown in the Appendix

3.2.1 Main Menu

- measurement menu,

- special functions menu,

- settings menu,
- calibration menu,
- saving of configuration parameters from the memory on the FLASH disk. Values
of parameters saved on the disk remain unchanged after device is switched off.
Actual values set during operation of the device reside in RAM which is
destroyed, after the device switch off. After power on of the device values of
configuration parameters from the FLASH disk are loaded.
- loading of configuration parameter values preset by manufacturer. This can be
used after extensive altering of configuration and unexpected behavior of the
device
3.2.2 Measurement Menu
- Range menu,
- Energy menu,
- operation mode selection menu with regard of wires number and power type.
Available modes are described in chapter 5.9.1 Operation Modes.
- setting of Time Base value
- setting the number of harmonics, which will be measured – FUNDAMENTAL
ONLY, FIRST n HARMONICS, or ALL 63 HARMONICS,
- setting the limit number - N of harmonics, which will be measured and evaluated.
This parameter also influences the number N from menu item „Harm. Meas
Mode". for the second option.
Note: If ALL 63 HARMONICS option was chosen in previous menu, then the
parameter in this menu is placed in parentheses which means that this option will
not be considered during measurements.
MEASURED or FROM P&Q (by calculation from active and reactive power)
3.2.3 Special Functions Menu
- displays the Vector Diagram i.e. phases of all 6 channels (3 x U, 3 x I)
RS-UG91-EN
Reference Standard RS 2x30, RS 1x30
...contrast enhancement in large steps
...contrast decrease in large steps
...contrast enhancement in small steps
...contrast decrease in small steps
User's Guide