System Menu; Calibration Menu; Calibration Menu „More Calib - Applied Precision RS 2130 User Manual

Reference standard, standard meter of electrical power and energy
Table of Contents

Advertisement

Harmonic Graph ->
History
Talk Only ...
Fout Response ...
Keyboard Test
About
RS-232...
Display....
Keyboard ...
Watchdog
Date & Time
Control Port
Remark: If the keypad is locked, a small icon with a figure of key appears in the top of display. To unlock
the keypad enter the password and press OK.
More Calib...
Per. Autocalib
Autocalib
Access to this menu is password protected.
Calib Funcs ...
Calib Const. ...
www.appliedp.com
- graphic representation of harmonic content. There is a chart on the left and
by up and down arrows zooming is changed in range 1x - 20x. On the right
side there is a list of harmonic items where the 1
means the number of harmonic item, 2
item's amplitude in % related to the fundamental harmonic and the 3
(ph[°]) states for phase shift of harmonic item. Paging of this list is possible via
left and right arrows on the keypad.
- Measured history values menu. This feature is described in chapter
5.16 Measured Values History System.
- settings of menu 'TALK ONLY'. Parameter value STATE is OFF or ON for
disabling resp. enabling of this mode. Item „COM..." sets this mode specifically
for this port in terms of its state, format and transferred quantities.
- setting the responses of impulse output. QUICK MODE – turning on the mode
of quick responses, parameter THRESHOLD holds the threshold value stated
in %.
- launching of keypad test

3.2.4 System Menu

- summary of basic device information: device type, device model, device serial
number, (device class), hardware version and firmware version.
- RS 232 menu
- display menu
- keyboard menu contains items for the setting of short voice signal generation
by each operation of keys (BEEP) and settings of keypad locking feature. Keypad
locking has these options: LOCK@SWITCH ON for locking the keypad right after
each device switch on, LOCK PASSWORD for password change and LOCK
NOW ->.
- parameter activating/deactivating the device watchdog. If it is allowed, then in
case of device freezing, the device is reset.
- actual device date and time, this is possible to change by editing,
- user output-input port menu control,

3.2.5 Calibration Menu

- maintenance calibration and calibration constants menu
This calibration menu is protected by a password – it is intended for change of
the device calibration constants or for the device calibration by the manufacturer
or other qualified person.
- parameter (ON) or (OFF) enabling or disabling automatic call of short device
selfcalibration
- short device selfcalibration activation (lasts for approx. 10 sec.). The short
calibration calibrates input circuits of device against internal DC voltage
reference. This helps to suppress the influence of ambient temperature changes.
3.2.6 Calibration Menu „More Calib..."
- setting of calibration functions
- setting of calibration constants
Reference Standard RS 2x30, RS 1x30
nd
column (U[%]) states for harmonic
RS-UG91-EN
User's Guide
st
column from left (Harm)
rd
column
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rs 1130Rs 1330Rs 2330

Table of Contents